摘要 患者男性,46岁,既往诊断为肥厚型心肌病,多源室性心动过速(VT),行"埋藏式心脏转复除颤器(ICD)植入术"。因"反复心慌、胸闷"再次就诊,行动态心电图检查,动态心电数据导录时发现仪器损坏。程控事件记录显示动态心电图检查期间发生快速VT,ICD予连续3次电击终止。经分析发现,仪器损坏系ICD多次高能量电击超过了动态心电图仪安全电流所致。
出处 《中国心脏起搏与心电生理杂志》 2021年第1期88-89,共2页 Chinese Journal of Cardiac Pacing and Electrophysiology
  • 相关文献

参考文献3

二级参考文献13

  • 1Daubert JP, Zareba W, Cannom DS, et al. Inappropriate implant- able cardioverter-defibrillator shocks in MADIT II: frequency, mechanisms, predictors and survival impact[ J]. J Am Coil Cardiol, 2008,51 : 1 357.
  • 2Sweeney MO, Wathen MS, Volosin K, et al. Appropriate and inap- propriate ventrieular therapies, quality of life, and mortality among primary and secondary prevention implantable cardioverter defibrilla- tor patients: results from the pacing fast VT reduces shock therapies ( PainFREE Rx Ⅱ) trial[ J ]. Circulation, 2005,111:2 898.
  • 3Pedersen SS, VAN DEN Berg M, Erdman RA, et al. Increased anxiety in partners of patients with a cardioverter-defibrillator: the role of indication for ICD therapy, shocks, and personality[ J]. Pa- cing Clin Electrophysiol, 2009,32 : 184.
  • 4Bilge AK, Ozben B, Demircan S, et al. Depression and anxiety sta- tus of patients with implantable cardioverter defibrillator and precipi- tating factors[J]. Pacing Clin Electrophysiol, 2009,32:184.
  • 5van Rees JB, Borleffs C J, de Bie MK, et al. Inappropriate implantable cardioverter-defibrillator shocks: incidence, predictors and im- pact on mortality[J]. J Am Coil Cardiol, 2011,57:556.
  • 6Gunderson BD, Abeyratne A1, Olson WH, et al. Effect of pro- grammed number of intervals to detect ventricular fibrillation on im- plantable cardioverter-defibrillator aborted and unnecessary shocks [ J ]. Pacing Clin Electrophysiol,2007,30 : 157.
  • 7Wilkoff BL, Williamson BD, Stern RS, et al. Strategic program- ming of detection and therapy parameters in implantable cardioverter- defibrillators reduces shocks in primary prevention patients: results from the PREPARE (Primary Prevention Parameters Evaluation ) study[J]. J Am Coil Cardiol,2008, 52:541.
  • 8Kuhlkamp V, Wilkoff BL, Brown AB, et al. Experience with adual chamber implantable defibrillator [J]. Pacing Clin Eleetrophysiol, 2002, 25 : 1 041.
  • 9Friedman PA, McClelland RL, Bamlet WR, et al. Dual-chamber versus single-chamber detection enhancements for implantable defib- rillator rhythm diagnosis: the detect supravenntricular tachycardia study[ J]. Circulation 2006, 113:2 871.
  • 10向晋涛,江洪.埋藏式心脏转复除颤器治疗的心理问题[J].中国心脏起搏与心电生理杂志,2009,23(1):15-18. 被引量:29