期刊文献+

口罩呼吸阻力检测仪流量校准关联性压力分析 被引量:3

Correlation Pressure Analysis for Flow Calibration of Mask Respiratory Resistance Tester
原文传递
导出
摘要 以ZR-1210型口罩呼吸阻力检测仪为校准对象,分别采用小孔微压与电子孔口流量计对仪器通气特定流量点85L/min进行校准,并对仪器关联性压力进行了分析:试验结果表明采用电子孔口流量计校准仪器通气流量,其所关联的仪器压力示值均不超过±1000Pa,在仪器安全性能方面较小孔微压流量法具有一定的优势同时,提出的流量校准中关联性压力分析为口罩呼吸阻力检测仪流量校准选择标准器提供了一种技术参考. In this paper,the ZR-1210 mask breathing resistance detector was used as the calibration object,and the small hole micro-pressure and electronic orifice flowmeter were used to calibrate the specific flow point of the instrument at85 L/min,and the correlation pressure of the instrument was analyzed. The test results show that the electronic orifice flowmeter which was used to calibrate the instrument was of certain advantages over the small-pore micro-pressure method in terms of instrument safety performance,as the associated pressure of the instrument is not more than ± 1000 Pa. At the same time,the proposed correlation pressure analysis in the flow calibration provides a technical reference for selecting a standard for the flow calibration of the mask respiratory resistance tester.
作者 李德林 李艾橘 胡启亮 温晓东 叶穗波 LI Delin;LI Aiju;HU Qiliang;WEN Xiaodong;YE Suibo(Dongguan Institute of Metrology Guangdong Academy of Metrology,Dongguan 523343,China)
出处 《计量科学与技术》 2021年第2期73-75,共3页 Metrology Science and Technology
关键词 口罩呼吸阻力检测仪 流量校准 压力分析 mask respiratory resistance tester flow calibration pressure analysis
  • 相关文献

参考文献6

二级参考文献30

  • 1孙建丽,沈恒根.自吸过滤式防微粒口罩过滤性能试验研究[J].产业用纺织品,2006,24(8):37-40. 被引量:8
  • 2GB2626--2006,呼吸防护用品自吸过滤式防颗粒物呼吸器[s].
  • 3路乘风;崔政斌.防尘防毒技术,2004.
  • 4佘启元.个人防护装备知识与标准实用全书,2002.
  • 5Gomersall C D, Tai DYH, Loo S, et al. Expanding ICU facilities in an epidemic: Recommendations based on experience from the SARS epidemic in Hong Kong and Singapore[J]. Intensive Care Medicine, 2006, 32 : 1 004 - 1 013.
  • 6Spies Adri, Kerry S Wilson, Ferrie Robert. Respirator fit of a medium mask on a group of south Africans: A cross-sec- tional study[J]. Environmental health, 2011, 10(1):1 -7.
  • 7R Kent Oestenstad, Leshan J Elliott, T Mark Beasley. The effect of gender and respirator brand on the association of res- pirator fit with facial dimensions[J]. Journal of Occupational and Environmental Hygiene ,2007,4(12) :923 -930.
  • 8Zhuang Z, Coffey C C, Ann R B. The effect of subject characteristics and respirator features on respirator rit[J]. Journal of Occupational and Environmental Hygiene, 2005,2(12) :641 -649.
  • 9Zhuang Z, Bradtmiller B, Shaffer R E. New respirator fit test panels representing the current U. S. civilian work force[J].Journal of Occupational and Environmental Hygiene, 2007, 4(9):647 -659.
  • 10程文娟,余丹,余艳艳,等.方形、杯状N95过滤型口罩适合性与头面部尺寸关系分析[c].第九届全国环境与职业医学研究生学术研讨会论文集,2010:45—48.

共引文献58

同被引文献48

引证文献3

二级引证文献6

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部