摘要 依据NiTi自膨式血管支架的工作机理及临床使用条件,研究不同热处理条件(包括温度和时间)后的NiTi自膨式血管支架在形状恢复能力的变化及耐腐蚀性能。分析不同热处理工艺条件下支架Af温度、弹性变形适应性、径向支撑力、局部抗挤压和耐腐蚀等性能指标的变化。结果表明:在热处理温度为500℃,热处理时间为10 min的试验条件下得到的NiTi自膨式血管支架的形状恢复能力为最佳状态,耐腐蚀性满足要求。 Effects of heat treatment conditions(including temperature and time)on the shape memory recovery and corrosion resistance of NiTi self-expanding vascular stents were studied based on working mechanism and clinical use.The Af temperature,dimensional recovery,crush resistance with radially applied load and point applied load of stents and corrosion resistance were characterized in diffident heat treatment conditions.The research results allow the conclusion that the stent treated at 500℃for 10 min has optimum performance,and corrosion resistance meets the requirements.
作者 王艳艳 罗健 李勇 陈洁云 汤京龙 吕原原 周俊蕾 WANG Yanyan;LUO Jian;LI Yong;CHEN Jieyun;TANG Jinglong;LYU Yuanyuan;ZHOU Junlei(Shanghai MicroPort Medical(Group)Co.Ltd.,Shanghai,201203;National Institute for Food and Drug Control,Beijing,100050)
出处 《中国医疗器械杂志》 2021年第2期119-124,共6页 Chinese Journal of Medical Instrumentation
基金 国家重点研发计划项目(2018YFC1106700)。
关键词 NiTi自膨式血管支架 形状恢复能力 热处理 耐腐蚀 NiTi self-expanding vascular stents shape memory recovery heat treatment corrosion resistance
  • 相关文献

参考文献5

二级参考文献87

  • 1赵振心,刘道志,张一.血管支架材料及其临床研究进展[J].中国医疗器械杂志,2005,29(6):391-395. 被引量:16
  • 2贺志荣,王芳,周敬恩.Ni含量和热处理对Ti-Ni形状记忆合金相变和形变行为的影响[J].金属热处理,2006,31(9):17-21. 被引量:40
  • 3贺志荣.Ti-Ni形状记忆合金多阶段可逆相变的类型及其演化过程[J].金属学报,2007,43(4):353-357. 被引量:24
  • 4张金山.严格掌握金属内支架置入术的适应证[J].中华放射学杂志,1998,32(10):653-654.
  • 5Lincoff AM, Furst JG, Ellis SG, et al. Sustained local delivery of dexamethasone by a novel intravascular eluting stent to prevent restenosis in the porcine coronary injury model. J Am Coll Cardiol, 1997, 29:808-816.
  • 6Tamai H, Igaki K, Kyo E, et al. Initial and 6-month results of biodegradable poly-l-lactic acid coronary stents in humans. Circulation, 2000, 102:399-404.
  • 7Vasenius J, Vainionpaa S, Vihtonen K, et al. Comparison of in vitro hydrolysis, subcutaneous and intramedullary implantation to evaluate the strength retention of absorbable osteosynthesis implants. Biomaterials, 1990, 11: 501- 504.
  • 8欧阳墉. 管腔内支架的作用机制和性能. 见:张金山,主编.现代腹部介入放射学. 北京:科学出版社,2000.239-241.
  • 9Stack TE, Califf RM, Phillips HR, et al. Interventional cardiac catheterization at Duke medical center. Am J Cardiol, 1988, 62(Suppl F): 3-24.
  • 10Fichman DL, Leon MB, Baim DS. A randomized comparison of coronary-stent placement and balloon angioplasty in the treatment of coronary artery disease. Stent Restenosis Study Investigators. N Engl Med, 1994, 331:496-501.