摘要
目的:利用双通道视觉质量分析系统(OQAS-II)评价泪膜稳定性对青少年视网膜成像质量的影响,为临床研究提供相关数据。方法:系列病例研究。选取2019年3─10月在福州市中医院眼科门诊就诊的青少年患者,按泪膜破裂时间(BUT)分为研究组(BUT<10 s)32例和对照组(BUT≥10 s)30例。利用基于双通道技术的OQAS-II系统检测2组患者自然状态下的调制传递函数截止频率(MTF cutoff)、客观散射指数(OSI)、不同对比度下的OQAS-II值(OV100%、OV20%、OV9%),同时检测2组患者在20 s持续睁眼过程中的OSI动态变化,获取此动态过程的OSI均值及标准差(SD-OSI)。采用独立样本t检验比较2组各数据间的差异性,采用Spearman秩相关分析法分析数据资料间的相关性。结果:2组患者的年龄、球镜度、柱镜度比较差异均无统计学意义。研究组的BUT为(5.90±2.16)s,明显小于对照组的(11.53±2.06)s,差异有统计学意义(t=-10.473,P<0.001)。2组MTF cutoff、OSI、OV100%、OV20%、OV9%值比较,差异无统计学意义;研究组OSI均值和SD-OSI高于对照组,差异有统计学意义(t=2.259,P=0.028;t=0.359,P=0.036)。MTF cutoff、OSI、OV100%、OV20%与BUT无相关性;而OV9%值与BUT呈正相关(r=0.273,P=0.032)。20 s内的OSI均值与BUT呈负相关(r=-0.276,P=0.002),SD-OSI与BUT无相关性。结论:青少年泪膜不稳定可引起视网膜成像质量下降。OQAS-II系统在评价青少年泪膜质量中具有很高的应用价值。
Objective:To evaluate the impact of tear film stability on adolescent visual quality by using the Optical Quality Analysis System(OQAS-II),and to provide relevant data for clinical research.Methods:In this case series study,32 adolescent patients with short break-up time(BUT<10 s)and thirty control subjects(BUT≥10 s)were enrolled from the outpatient clinic of Fuzhou Hospital of Traditional Chinese Medicine from March to October 2019.The OQAS-II system was used to detect the modulation transfer function cut off frequency(MTF cutoff),objective scattering index(OSI)and OQAS values with contrasts of 100%,20%and 9%(OV100%,OV20%and OV9%)in the natural state.Meanwhile,the dynamic changes of OSI were detected during 20-second intervals with the eye open continuously,and the mean value and standard deviation of OSI(SD-OSI)in this dynamic process were obtained.Differences in the parameters were compared between the two groups using independent t test.Spearman rank correlation analysis was used to analyze the correlation between the data.Results:There were no statistically significant differences between the two groups in terms of age,spherical degree or astigmatic degree.BUT was 5.90±2.16 s in the study group,which was significantly less than that in the control group(11.53±2.06 s).This difference(t=-10.473,P<0.001)was statistically significant.MTF cutoff,OSI,OV100%,OV20%,OV9%were compared between the 2 groups,and the differences were not statistically significant.The mean values of OSI and SD-OSI in the study group were higher than those in the control group,and the difference was statistically significant(t=2.259,P=0.028;t=0.359,P=0.036).There was no correlation between MTF cutoff,OSI,OV100%,OV20%and BUT(all P>0.05)but the value of OV9%was positively correlated with BUT(r=0.273,P=0.032).The mean value of OSI in 20-second period was negatively correlated with BUT(r=-0.276,P=0.002),while SD-OSI was not correlated with BUT.Conclusions:The instability of adolescent tear films can lead to a decline in retinal image quality,and it is related to the degree of BUT.There is a high value of OQAS-II in evaluation of adolescent tear films.
作者
程慧勤
王段
伊良燕
宋曼
Huiqin Cheng;Duan Wang;Liangyan Yi;Man Song(Department of Ophthalmology,Fuzhou Hospital of Traditional Chinese Medicine,Fuzhou 350001,China)
出处
《中华眼视光学与视觉科学杂志》
CAS
CSCD
2021年第2期116-119,共4页
Chinese Journal Of Optometry Ophthalmology And Visual Science
基金
福州市科技计划项目(2018-S-102-6)。