摘要 本研究对冷原子荧光便携式痕量汞水质在线监测仪与两种实验室汞测定仪器进行了检出限、测定上限、准确度、精密度、离子干扰、实际水样加标回收率等指标的比较。这两种实验室汞分析仪是DMA-80痕量汞分析仪和AFS-230E双道原子荧光光度计。实验证明,集前处理于一体的冷原子荧光便携式痕量汞水质在线监测仪,性能上分析灵敏度高,准确度、精密度和抗离子干扰能力与两种实验室仪器无明显差异,能满足地表水汞在线监测的要求。 In the study,the detection limit,determination range,maximum quantitative detection limit,accuracy,precision,interference of ions and recovery rate of cold atomic fluorescence portable water quality automatic monitor for trace mercury,DMA-80 trace mercury analyzer and AFS-230E dual-channel atomic fluorescence spectrophotometer are compared.The experimental results show that the cold atomic fluorescence portable trace mercury water quality automatic monitor which integrated with pretreatment has higher analysis sensitivity,while its accuracy,precision and anti-ion interference ability are not significantly different from the two laboratory instruments,and it can meet the requirements of online mercury monitoring in surface water.
作者 李琴 施择 赵祖军 聂晶晶 黄海萍 王思苗 李雁 Li Qin;Shi Ze;Zhao Zujun;Nie Jingjing;Huang Haiping;Wang Simiao;Li Yan(YunNan Environmental Monitoring Center,Kunming 650034,Yunnan,China;Lihe Technology(Hunan)Co.,Ltd.,Changsha 410205,Hunan,China;Anning City Environmental Monitoring Station,Anning 650300,Yunnan,China)
出处 《分析仪器》 CAS 2021年第2期79-81,共3页 Analytical Instrumentation
基金 国家重大科学仪器设备开发专项“重点防控重金属汞、铬、铅、镉、砷便携/车载/在线监测仪器开发与应用示范”(2012YQ060115)资助项目。
关键词 冷原子荧光 便携式痕量汞水质在线监测仪 实验室汞测定仪 比较研究 Cold atomic fluorescence Portable trace mercury water quality on-line monitor Laboratory mercury analyzer Comparative study
  • 相关文献

同被引文献19

引证文献2

二级引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部