期刊文献+

医用导管亲水涂层附着力研究 被引量:3

Research on Adhesion of Hydrophilic Coating on Medical Catheters
下载PDF
导出
摘要 考察自制的一次性超滑硅胶导尿管、一次性超滑乳胶导尿管、亲水球囊扩张导管亲水涂层的附着力。方法是采用摩擦系数测试仪分别对3种医用导管未涂涂层、有涂层及浸泡后管体反复测试30次摩擦系数,并对测试后的管体进行染色。结果表明,相较于未涂涂层管体,亲水涂层导管摩擦系数降低了90%以上;30次摩擦系数测试后涂层摩擦系数均<0.1,且U2/U1≤1,测试后的涂层可被均匀染色。3种自制导管亲水涂层的附着力良好。 The adhesion of hydrophilic coating on self-prepared hyper-slippery silicone catheter,hyper-slippery latex catheter and hydrophilic balloon dilation catheter is tested.The friction coefficient of uncoated,coated and immersed tubes are tested by friction coefficient tester for 30 times,and then they are stained with Congo red.The results show that the friction coefficient of coated tubes is decreased by 90%as compared to uncoated tubes.After chafing for 30 times,the friction coefficient of the coating is less than 0.1.The coating can be dyed uniformly and U2/U1≤1.It is revealed that the adhesion of the three self-prepared hydrophilic catheters are excellent.
作者 刘秋华 夏凯丽 牛利卫 王富利 崔景强 程杰 刘志远 师晓亭 张铭 Liu Qiuhua;Xia Kaili;Niu Liwei;Wang Fuli;Cui Jingqiang;Cheng Jie;Liu Zhiyuan;Shi Xiaoting;Zhang Ming(Henan Tuoren Medical Device Co.,Ltd.,Henan 453400,China;Key Laboratory of Henan Medical Polymer Materials Technology and Application,Xinxiang,Henan 453400,China)
出处 《涂层与防护》 2021年第3期50-54,共5页 Coating and Protection
关键词 医用导管 亲水涂层 附着力 摩擦系数 medical catheter hydrophilic coating adhesion friction coefficient
  • 相关文献

参考文献3

二级参考文献15

  • 1王飞,司俊杰.镀膜瓶塞[J].世界橡胶工业,2003,30(6):20-20. 被引量:1
  • 2张承焱.漫谈医用聚氨酯[J].世界橡胶工业,2003,30(5):45-48. 被引量:15
  • 3Langer R, Vacanti J P. Science, 1993, 260(4) : 920 ~ 926.
  • 4Puleo D A, Holleran L A, Doremus R H, et al. J Biomed Mat Res, 1991, 23:711 ~ 723.
  • 5Belanger M C, Marois Y, Roy R, et al. Artifi Org, 2000, 24( 11 ) : 879 ~ 888.
  • 6Ratcliffe A. Matr Bio, 2000,19 (4) : 353 ~ 357.
  • 7Weston M W, Laborde D V, Yoganathan A P. A Biomed Eng, 1999, 27(4) : 527 ~ 579.
  • 8Sheikh N, Katbab A, Mirzadeh H. Int J Adhesion & Adhesives,2000,20(4) : 299 ~ 304.
  • 9Skarja G A, Woodhouse K A. J Appl Polym Sci, 2000, 75 : 1522 ~ 1534.
  • 10姜春鹏,常津,许晓秋,等.亲水性聚合物修饰医疗器械表面的发展[J].化学通报,1999,(12):36-40.