AccuContour软件自动勾画镜面人患者4D-CT图像心脏结构的应用分析 被引量:2

Application Analysis of the AccuContour Software in Automatic Delineating the Structure of Heart with the 4D-CT images for Mirror People
下载PDF
导出
摘要 目的评估AccuContour软件自动勾画镜面人患者四维CT(4D-CT)图像心脏结构的可行性与准确性。方法选取1例镜面人肺癌患者的10套4D-CT图像序列进行研究,10套图像序列对应10个呼吸时相。对10套CT序列,使用Eclipse治疗计划系统手工勾画心脏结构;使用AccuContour软件自动勾画心脏结构;使用WPS表格,对自动勾画和手工勾画的体积差异、位置差异、形状一致性和勾画时间进行统计分析。结果自动和手工勾画两种方式心脏结构的体积差异(△V%)为(5.63±2.80)%;总位置差异(ΔL)为(2.28±0.45)mm;Dice相似性系数、敏感性指数、包容性指数和Jaccard系数的值分别为0.909±0.011、0.936±0.012、0.884±0.020和0.835±0.019;自动勾画时间平均约为49 s。结论使用基于深度学习的AccuContour软件自动勾画镜面人患者4D-CT图像的心脏结构,能够达到比较满意的勾画结果,使用该软件可明显缩短勾画时间,有效提高工作效率。 Objective To evaluate the feasibility and accuracy of the AccuContour software in automatic delineating the heart structure with 4D-CT images for mirror people.Methods Ten sets of 4D-CT image series of mirror people with lung cancer were studied.Ten sets of CT series were corresponding to ten respiratory phases.For the ten sets of CT series,the heart structures were manually delineated in the treatment planning system of Eclipse,and automatically delineated by the AccuContour software.The WPS sheet was used to statistically analyze the differences in volume,position,shape,and delineation time between the automatic and manual delineations.Results The differences in volume(ΔV%)between the two methods was(5.63±2.80)%;the total position difference(ΔL)was(2.28±0.45)mm;the values of dice similarity coefficient,sensitivity index,inclusiveness index and Jaccard coefficient were 0.909±0.011,0.936±0.012,0.884±0.020 and 0.835±0.019,respectively;the average time of automatic delineation was about 49 s.Conclusion The AccuContour software based on deep learning could achieve satisfactory results of automatic delineation of the heart structure of 4D-CT images for the mirror people.By using the software,it could significantly shorten the delineation time and effectively improve the work efficiency.
作者 时飞跃 秦伟 赵紫婷 王敏 赵环宇 魏晓为 SHI Feiyue;QIN Wei;ZHAO Ziting;WANG Min;ZHAO Huanyu;WEI Xiaowei(Radiation Therapy Center,Nanjing First Hospital,Nanjing Medical University,Nanjing Jiangsu 210006,China;Department of Medical Equipment,Nanjing First Hospital,Nanjing Medical University,Nanjing Jiangsu 210006,China;Center of Medical Physics,Nanjing Medical University,Nanjing Jiangsu 210029,China)
出处 《中国医疗设备》 2021年第4期1-4,20,共5页 China Medical Devices
基金 国家自然科学基金面上项目(81773240) 江苏省自然科学基金面上项目(BK20181118)。
关键词 AccuContour软件 自动勾画 心脏结构 镜面人 四维CT AccuContour software automatic delineation heart structure mirror people 4D-CT
  • 相关文献

参考文献13

二级参考文献59

共引文献74

同被引文献14

引证文献2

二级引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部