期刊文献+

基于温度敏感聚合物PMEO_(2)MA和GO复合纳米材料修饰传感膜测定肾上腺素

Determination of adrenaline based on temperature-sensitive polymer PMEO_(2)MA and GO nano-composite modified sensing membrane
下载PDF
导出
摘要 制备了一种以2-甲基-2-丙烯酸-2-(2-甲氧基乙氧基)乙酯(PMEO_(2)MA)和氧化石墨烯(GO)修饰在玻碳电极(GCE)上的温度响应型复合传感器,实现了肾上腺素电化学行为的温度调控.当温度低于聚合物的最低临界胶体温度(LCST)时,PMEO_(2)MA链处在拉伸状态,GO层与层之间距离变大,传感器导电性变差,从而不能检测到肾上腺素的电化学信号;当温度高于聚合物LCST时,PMEO_(2)MA链处在紧缩状态,GO层与层之间相互靠近,修饰膜电子传递速率变快,肾上腺素的电化学行为开启.研究结果表明:在聚合物LCST之上,复合传感膜的线性范围表现出0~45μM的线性响应,检出限是1.35μM;用于肾上腺素注射液样品检测时,其回收率为99.0%~103.9%. A glassy carbon electrode(GCE)modified by poly(2-methyl-2-acrylic acid-2-(2-methoxyethoxy)ethyl)(PMEO_(2)MA)and graphene oxide(GO)is prepared,and the formed temperature-responsive composite sensor realizes the temperature regulation of the electrochemical behavior of adrenaline.When the temperature is below the lowest critical colloidal temperature(LCST)of the polymer,the PMEO_(2)MA chain is in a stretched state,the distance between the GO layers gets larger,and the sensor conductivity becomes weaker,so that the electrochemical signal of adrenaline cannot be detected.When the temperature is above the polymer LCST,the PMEO_(2)MA chain is in a compressed state,the GO layers get closer to each other,the electron transfer rate of the modified membrane is increased,and the electrochemical behavior of adrenaline is turned on.The results of the study show that the linear range of the composite sensing membrane is 0~45μM above the LCST,and the detection limit is 1.35μM.When it is used for the detection of adrenaline injection samples,the recovery rate is 99.0%~103.9%.
作者 陈超 陈思 王晨曦 赵鹏程 彭天英 CHEN Chao;CHEN Si;WANG Chenxi;ZHAO Pengcheng;PENG Tianying(School of Materials and Chemical Engineering,Hunan City University,Yiyang,Hunan 413000,China;Xiangtan Medicine&Health Vocational College,Xiangtan,Hunan 411100,China;School of Chemistry,Xiangtan University,Xiangtan,Hunan 411100,China)
出处 《湖南城市学院学报(自然科学版)》 CAS 2021年第2期73-78,共6页 Journal of Hunan City University:Natural Science
基金 国家自然科学基金项目(21874114)。
关键词 温度响应型聚合物 氧化石墨烯 肾上腺素 检出限 temperature-responsive polymer graphene oxide adrenaline detection limit
  • 相关文献

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部