期刊文献+

三段控温固定点炉复现铝凝固点研究 被引量:6

Study on the Realization of Aluminum Freezing Point in the Three-zone Temperature Controlled Fixed Point Furnace
下载PDF
导出
摘要 固定点炉垂直温场均匀性是影响ITS-90国际温标固定点温坪质量的重要因素。为了提高铝凝固温坪的复现水平,设计了三段控温固定点炉,利用金属外壳铝固定点容器,研究了垂直温场均匀性及其影响因素;在此基础上,采用连续热流密度法高精度复现了铝凝固点。实验结果表明:通过调整上部、中部、下部炉温的设置,可改善固定点炉温场均匀性,铝凝固温坪4 h内的温度变化在1 m K以内;此外,在实验条件下,高纯铝相变的过冷度约为0.14℃。 The vertical temperature uniformity of the fixed-point furnace is an important factor affecting the plateau quality of the fixed-points of the International Temperature Scale of 1990( ITS-90). In order to improve the reproducibility of the aluminum freezing point,a three-zone temperature-controlled fixed-point furnace was designed,and the aluminum fixed point cell with a metal shell was used to study the vertical temperature uniformity and its influencing factors. On this basis,the continuous heat flux method was adopted to precisely realize the aluminum freezing point. The experimental results show that the temperature uniformity can be improved by adjusting the settings of the top,middle,and bottom furnace temperatures,and the temperature stability of the freezing plateau within 4 hours is less than 1 mK. The phase transition supercooling of the high purity aluminum is approximately 0. 14 ℃ under the experimental conditions.
作者 何沛 林林 闫小克 张明宇 王宁 HE Pei;LIN Lin;YAN Xiao-ke;ZHANG Ming-yu;WANG Ning(University of Science and Technology Beijing,Beijing 100083,China;National Institute of Metrology,Beijing 100029,China)
出处 《计量学报》 CSCD 北大核心 2021年第3期321-326,共6页 Acta Metrologica Sinica
基金 国家重点研发计划项目(2017YFF0205901)。
关键词 计量学 铝凝固点 三段控温 固定点炉 垂直温场 凝固温坪 ITS-90 metrology aluminum freezing point three-zone temperature controlled fixed point furnace vertical temperature uniformity freezing plateau ITS-90
  • 相关文献

参考文献9

二级参考文献38

共引文献19

同被引文献37

引证文献6

二级引证文献9

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部