不同机械指数超声对微泡的作用 被引量:4

Effect of ultrasound with different mechanical indexes on microbubbles
下载PDF
导出
摘要 目的体外观察不同机械指数超声对微泡造影剂的作用。方法设置不同机械指数,采集超声体外辐照纤维素管内微泡的造影图像,观察其中微泡浓度变化,分析不同机械指数条件下微泡的运动趋势;拟合微泡运动速率曲线,分析微泡运动规律。结果低机械指数对微泡无明显破坏作用,微泡运动速率呈线性变化;随着机械指数增加,微泡运动速率增加;机械指数>0.22时,微泡在超声作用下迅速向对侧以较高速率运动,运动至对侧后出现破裂现象。结论不同机械指数下,超声对微泡造影剂的作用存在区别,低机械指数时以推移作用为主,高机械指数可致微泡破裂。 Objective To observe the in vitro effect of ultrasound with different mechanical indexes on microbubbles contrast agent.Methods Different mechanical indexes were set to acquire the contrast images of microbubbles in ultrasonic irradiation cellulose tubes.Changes in the concentration of microbubbles in the images were observed,and the movement trend of microbubbles under different mechanical indexes were analyze.Then the velocity curve of microbubbles was fitted,and movement of microbubbles were estimated.Results The microbubbles were not damaged at low mechanical indexes,and the movement rate changed linearly.The movement rate of microbubbles increased with the increase of mechanical indexes.At the mechanical indexes higher than 0.22,microbubbles moved rapidly to the opposite side with a relatively high rate under the action of ultrasound,and burst after moving to the opposite side.Conclusion The effect of ultrasound on microbubbles contrast agent was different under different mechanical indexes.Microbubbles were displaced at low mechanical indexes,and were distinctly destroyed at high mechanical indexes.
作者 位红芹 付伸伸 马穗红 邱春花 柳建华 WEI Hongqin;FU Shenshen;MA Suihong;QIU Chunhua;LIU Jianhua(Department of Ultrasound Medicine,the Second Affiliated Hospital of South China University of Technology,Guangzhou First People's Hospital,Guangzhou 510180,China)
出处 《中国医学影像技术》 CSCD 北大核心 2021年第5期648-651,共4页 Chinese Journal of Medical Imaging Technology
关键词 机械指数 微泡 超声检查 造影剂 mechanical indexes microbubbles ultrasonography contrast media
  • 相关文献

参考文献2

二级参考文献17

  • 1赵玉英,智光,徐勇,崔文瑾,王身国,何振山.新型聚内酯超声造影剂的体外基本特性研究[J].中华超声影像学杂志,2006,15(5):381-383. 被引量:5
  • 2Tlaxca JL, Rychak J J, Ernst PB, et al. Ultrasound-based molec- ular imaging and specific gene delivery to meseBteric vaseulature by endothelial adhesion molecule targeted microbubbles in a mouse model of Crohn's disease. J Control Release, 2013, 165 (3) :216-225.
  • 3Ren ST, Liao YR, Kang XN, et al. The antitumor effect of a new doeetaxel loaded microbubble combined with low-frequency ultrasound in vitro: Preparation and parameter analysis. Pharm Res, 2013,30(6) :1574-1585.
  • 4Wang YH, Liao AH, Chen JH, et al. Photoacoustic/ultrasound dual-modality contrast agent and its application to thermotherapy. J Biomed Opt, 2012,17(4):045001.
  • 5Browning RJ, Mulvana H, Tang MX, et al. Effect of albumin and dextrose concentration on ultrasound and microbubble media-ted gene transfection in vivo. Ultrasound Med Biol, 2012,38(6): 1067- 1077.
  • 6Correas JM, Quay SD. EchoGen emulsion: A new ultrasound contrast agent based on phase shift colloids. Clin Radiol 1996,51 (Suppl 1) : 11-14.
  • 7Albrecht T, Cosgrove DO, Correas JM, et al. Renal, hepatic, and cardiac enhancement on Doppler and gray-scale sonograms obtained with EchoGen. Acad Radiol, 1996, 3 (Suppl 2) : 198- 200.
  • 8钟华,李锐,郝迎学,张萍,郭燕丽,薛雅芳,卓丽莎.超声空化效应对肿瘤细胞迁移运动影响的实验研究[J].中国超声医学杂志,2008,24(2):97-99. 被引量:3
  • 9万鲲,高申,费翔,张顺欣.高分子聚合物PLGE超声微泡的制备与体外显影研究[J].中国药物应用与监测,2010,7(6):335-337. 被引量:2
  • 10张群霞,骆杰,冉海涛,王志刚,郑元义,张辉,钟世根,李巧.比较脂质微泡与聚合物超声造影剂的微循环流变学[J].中国医学影像技术,2011,27(8):1525-1527. 被引量:6

共引文献7

同被引文献21

引证文献4

二级引证文献6

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部