摘要
建立了测定大气中25种有机氯农药(OCPs)的同位素稀释-高分辨气相色谱/高分辨质谱法(ID-HRGC/HRMS)。样品用正己烷/二氯甲烷(1∶1,v/v)进行加速溶剂萃取(ASE)。通过柱洗脱实验、单柱和组合柱净化实验,最终确定样品的净化方案为弗罗里硅土固相萃取柱和石墨化炭黑固相萃取柱组合净化。样品萃取液净化后进行HRGC/HRMS分析。采用平均相对响应因子(RRF)法对样品中目标物进行定量,6点校准溶液RRF的相对标准偏差(RSD)均≤20%。线性范围为0.4~800μg/L,相关系数R2均>0.992。对空白样品依次进行100 pg、400 pg和15 ng水平下的加标试验,各添加水平下OCPs测定值的RSD为0.64%~16%,加标回收率为67.2%~135%。穿透试验表明,滤膜+聚氨酯泡沫/聚氨酯泡沫作为吸附介质的大体积主动大气采样器(AAS)在采集环境空气时,五氯苯极易发生穿透,有效采样模式待进一步研究。在上述采样模式下,六氯苯的有效采样体积较小,标准状态(101.325 kPa,273 K)采样体积应≤30 m3,其他OCPs应≤1200 m3。以上述体积计算,25种目标化合物的检出限为0.002~0.7 pg/m3。对北京环境空气样品分析测定,结果显示除反式-环氧七氯、异狄氏剂、顺式-九氯和4,4′-滴滴滴在部分样品中未检出外,其他OCPs均为100%检出;六氯苯浓度在514~563 pg/m3之间,其他OCPs的浓度在0.01~18.9 pg/m3之间;替代标回收率为33.9%~155%。由于现有相关监测标准的仪器灵敏度较低、方法检出限较高,已无法满足目前空气中痕量OCPs的测定需求,因此亟待修订新的高灵敏度监测方法标准。该方法适用于目前大气中OCPs的超痕量水平分析,为新标准的制订奠定基础,也为国家履行相关国际公约提供有力技术指导。
A method for the determination of 25 organochlorine pesticides(OCPs)in the atmosphere using isotope dilution high-resolution gas chromatography/high-resolution mass spectrometry(ID-HRGC/HRMS)was developed.Sample extraction was performed using an accelerated solvent extractor(ASE).The extraction parameters were as follows:the extraction solvent was 50%(v/v)hexane in dichloromethane,the extraction temperature was 100℃,the static time was 8 min,the cell was rinsed with 60%cell volume using the aforementioned extraction solvent,the purging time was 180 s with N2 gas,and the extraction proceeded through three cycles.The eluting solutions of common cartridges such as florisil,graphitized carbon black,alumina,and silica were determined via cartridge elution tests.Use of the aforementioned cartridges alone cannot remove the pigments in the air sample solution.Subsequently,all possible pairwise combinations of the four cartridges were used for sample cleaning,and only the combination of florisil and graphitized carbon black was found to completely remove the pigments.Thus,the combination of florisil and graphitized carbon black cartridges using 10 mL toluene for elution was determined as the final cleaning method in this study.A high-resolution mass spectrometer equipped with a gas chromatograph was used for quantification.A fused-silica capillary column(Rtx-CL Pesticides2,30 m×0.25 mm×0.2μm)was used to separate the target compounds.Injection was performed in the splitless mode at 250℃.The flow rate of nitrogen gas was maintained constant at 1 mL/min.The oven temperature was 110℃(1 min),20℃/min up to 210℃,1.5℃/min up to 218℃(1 min),and 2℃/min up to 260℃(1 min).HRMS was conducted at>8000 resolution,the source temperature was 280℃in the electron impact mode using ionization energy of 35 eV,and measurements were performed in the selective ion monitoring(SIM)mode.Twenty-five OCPs were identified by comparing their GC retention times with those of the corresponding labeled compounds,and the actual ion abundance ratios of two exact m/z values with the corresponding theoretical values.The 25 OCPs were quantified by average relative response factors(RRFs),and the relative standard deviations(RSDs)of the RRFs with six calibration solutions were no more than 20%.The linear range of this method was 0.4 to 800μg/L,and the correlation coefficients(R2)were higher than 0.992.To validate the method,clean materials(one quartz fiber filter(QFF)and two polyurethane foam(PUF)plugs)were spiked with 100 pg,400 pg,and 15 ng native OCP standards,respectively;the RSDs of the 25 OCPs for each spiked level ranged from 0.64%to 16%.The spiking recoveries of the native OCPs ranged from 67.2%to 135%.Penetration experiments were conducted by sampling various volumes of air(15-1000 m3)using a filter-PUF/PUF high-volume active sampler.The breakthrough volume was sampled when the amount of OCPs collected in the PUF of the non-sampling end reached 5%of the total amount collected by both PUFs.When a high-volume active sampler with filter-PUF/PUF was used as an adsorbent for sampling atmospheric OCPs,a serious breakthrough of pentachlorobenzene(PeCB)occurred.The effective sampling volume of hexachlorobenzene(HCB)was very low,and was no more than 30 m3 under the standard conditions(101.325 kPa,273 K).The effective sampling volumes of other OCP compounds should be no more than 1200 m3.This will necessitate the use of high-adsorption-capacity adsorbents such as the PUF-XAD(a styrene-divinylbenzene copolymer)sandwich used for sampling air PeCB and HCB.Calculation with the effective sampling volumes from the penetration experiment revealed that the limits of detection of the 25 OCPs were in the range of 0.002 to 0.7 pg/m3.Thus,the detection levels of OCPs in this study were reduced to at least 2%of the current monitoring standards.Analysis of air samples in Beijing showed that all the target compounds except for trans-heptachlor epoxide,endrin,cis-nonachlor and 4,4′-DDD were 100%detected in the air samples.The concentrations of HCB(in volumes of 15-30 m3)ranged from 514 to 563 pg/m3,while those of the other OCPs(in a volume of 600 m3)ranged from 0.01 to 18.9 pg/m3.The recoveries of surrogate standards in this sample analysis were in the range of 33.9%to 155%,which satisfied the requirements of EPA Method 1699.Because of the very high detection limits,the current related monitoring standards cannot meet the requirements of atmospheric OCP analysis,especially at the ultra-trace level.In addition,highly sensitive monitoring standard methods are urgently needed.This method is suitable for analyzing most atmospheric OCPs,even at the ultra-trace level.It also lays the foundation for a new standard method formulation and provides strong support for the implementation of relevant international conventions.
作者
张静星
郑晓燕
谭丽
刘进斌
于海斌
ZHANG Jingxing;ZHENG Xiaoyan;TAN Li;LIU Jinbin;YU Haibin(China National Environmental Monitoring Centre, Beijing 100012, China)
出处
《色谱》
CAS
CSCD
北大核心
2021年第5期541-551,共11页
Chinese Journal of Chromatography
基金
国家自然科学基金项目(41101476).
关键词
同位素稀释法
高分辨气相色谱/高分辨质谱
有机氯农药
环境空气
isotope dilution(ID)
high-resolution gas chromatography/high-resolution mass spectrometry(HRGC/HRMS)
organochlorine pesticides
ambient air