摘要
A new pressure-calibration method for calibrating the reduction of second harmonic(2f)amplitude caused by pressure broadening effect in sealed microbial growth environment is present.The new method combines with linewidth compensation and modulation depth compensation and makes the 2f amplitude accurately retrieve metabolic CO_(2) in microbial growth.In order to verify the method,a simulation experiment is developed,in which the increasing CO_(2) concentration leads to the increasing pressure.Comparing with the relation between the traditional 2f amplitude and gas concentration,there is a monotonous relation between the calibrated 2f amplitude and CO_(2) concentration,particularly,a linear relation is present when the CO_(2) concentration is replaced with the CO_(2) particle number.In terms of microbial measurement,the growth of Escherichia coli is measured,and the culture bottle is sealed during the microbial growth process.The experimental results show that,comparing to the microbial growth retrieved by traditional 2f amplitude,the calibrated 2f amplitude can accurately retrieve microbial growth in sealed environment.
基金
the National Key Research and Development Program of China(Grant Nos.2017YFC0209700 and 2016YFC0303900)
the National Natural Science Foundation of China(Grant No.41730103).