摘要 [Objectives]To establish a UPLC-UV method for the determination of polydatin in the serum of Wistar rats.[Methods]Acquity UPLC BEH shield RP18 column(1.7μm,2.1 mm×100 mm,Waters Corporation,USA)was used as the analytical column,acetonitrile-water(55∶45)was used as the mobile phase,the flow rate was 0.5 mL/min,and the column temperature was 30℃and the detection wavelength was 306 nm.[Results]The linear range of the established serum sample analysis method was 1.0-20.0μg/mL,and the correlation coefficient was r=0.9994;the intraday and interday RSD of Wistar rat serum was less than 3.0%,and the accuracy was higher than 90%.[Conclusions]This method is sensitive,accurate,and rapid.It is suitable for monitoring the concentration of polydatin in serum after intragastric administration,and can also be used for pharmacokinetics and bioavailability studies.
出处 《Medicinal Plant》 CAS 2021年第2期63-64,69,共3页 药用植物:英文版
基金 Project of Traditional Chinese Medicine Administration of Guangxi Zhuang Autonomous Region(GZZC2019147) Project for Improving Basic Research Ability of Middle Aged and Young Teachers in Colleges and Universities of Guangxi in 2020(2020KY13034) The First Batch High-level Talent Scientific Research Project of The Affiliated Hospital of Youjiang Medical University for Nationalities in 2019(Y20196311) National Traditional Chinese Medicine Characteristic Technology Heritage Talent Training Program(2015481601003) Program of Youjiang Medical University for Nationalities(yy2018ky018).