摘要
若处于量热计中的反应室所产生的热量未能被完全测量,则会在用氮气作为标定气体,测量氢(氘)气与材料之间的反应热时,就会因标定和触发实验中所使用气体的导热系数的差异,测量到错误的过剩热功率。随着输入功率的增加,这种"过热"功率也会随之增大。通过使用稳定的热流式(塞贝克)量热计来讨论这种测量误差,并提出了减少误差的方法。同时对Rossi专利内容的真实性提出了质疑。
Measurement of a reaction between hydrogen(deuterium)gas and material using a calorimeter which is calibrated by nitrogen will show erroneous excess heat power production due to the difference of thermal conductivity of the gases used in the calibration and triggering experiments,which leads to the energy generated in the chamber of the calorimeter has not been totally measured.With the increase of input power,the excess power will increase.This insidious error is discussed by using a stable heat-flow(Seebeck)calorimeter and procedures to reduce error are put forward.And the authenticity of Rossi’s patent is questioned.
作者
王兴业
申炳俊
金丽虹
周丹
田坚
WANG Xing-ye;SHEN Bing-jun;JIN Li-hong;ZHOU Dan;TIAN Jian(School of Materials Science and Engineering,Changchun University of Science and Technology,Changchun 130022;School of Life Science and Technology,Changchun University of Science and Technology,Changchun 130022;School of Science,Changchun University of Science and Technology,Changchun 130022;Institute of Space Photoelectric Technology,Changchun University of Science and Technology,Changchun 130022)