基于三维预测剂量的自调节调强放疗自动计划方法 被引量:1

Self-adjustable automatic planning method of intensity modulated radiotherapy based on 3D predicted dose
原文传递
导出
摘要 目的开发一种基于预测剂量的自调节调强放射治疗自动计划方法,以增强自动计划的鲁棒性。方法利用3D U-Res-Net_B网络预测出三维剂量分布后,在直接子野优化的每次迭代中先基于上次迭代结果计算当前剂量,再联合预测剂量计算目标剂量,然后以此为目标进行优化。完成所有迭代或满足循环退出条件后,得到最终的治疗计划。在30例直肠癌病例上进行测试,验证算法的效果。结果临床计划治疗靶区的V100%均值和标准差为(95.03±0.91)%,自动计划为(94.67±1.96)%,接近临床值(P>0.05),而预测值为(92.90±2.13)%,与临床计划的差异具有统计学意义(t=29.0,P<0.05);自动计划在小肠V35、膀胱V40、股骨头的V20~V40等多项指标上低于预测值和临床值,且差异具有统计学意义(t=4.5~118.0,P<0.05),在其他危及器官的指标上与临床值的差异无统计学意义(P>0.05)。结论本方法增强了自动计划的鲁棒性,提高了其应对复杂情况的能力。 Objective To develope a self-adjustable automatic planning method of intensity modulated radiotherapy based on predicted dose,in order to enhance the robustness of automatic planning.Methods After the patients'dose by 3D U-Res-Net_B network was predicted,the current dose was calculated based on the last iteration result,then the predicted dose was combined to calculate the target dose and optimized.With all iterations completed or exit conditions satisfied,final treatment plannings would be acquired.A total of 30 cases of rectal cancer were tested to verify the effectiveness of the algorithm.Results The mean value of planning target volumes'V100%was(95.03±0.91)%for clinical plans,close to(94.67±1.96)%for automatical plans(P>0.05),and better than(92.90±2.13)%for predicted dose with the statisically significant difference(t=29.0,P<0.05).Automatic planning's indexes such as V35 of small intestines,V40 of bladders and V20-V40 of femoral heads were lower than predicted and clinical ones,with the statisically significant difference(t=4.5-118.0,P<0.05).Discrepancy in other indexes of organs at risk was not statistically significantly different(P>0.05).Conclusions This method made automatic planning processes more robust and more adaptive to difficult clinical situations.
作者 闫永恒 潘茂云 周解平 吴爱东 吴文华 徐榭 裴曦 Yan Yongheng;Pan Maoyun;Zhou Jieping;Wu Aidong;Wu Wenhua;Xu Xie;Pei Xi(School of Nuclear Science and Technology,University of Science and Technology of China,Hefei 230025,China;Anhui Wisedom Technology Co.Ltd,Hefei 230088,China;Department of Radiotherapy,Anhui Cancer Hospital,Hefei 230031,China;Department of Radiotherapy,The First Affiliated Hospital,University of Science and Technology of China,Hefei 230031,China;Anhui Drug Evaluation and Inspection Center,Anhui Drug Inspection Center,Hefei 230000,China)
出处 《中华放射医学与防护杂志》 CAS CSCD 北大核心 2021年第6期444-449,共6页 Chinese Journal of Radiological Medicine and Protection
基金 安徽省自然科学基金(1908085MA27) 安徽省重点研究与开发计划项目(1804a09020039)。
关键词 调强放射治疗 自动计划 自调节 Intensity modulated radiotherapy Automatic radiotherapy planning Self-adjustable
  • 相关文献

参考文献2

二级参考文献12

共引文献7

同被引文献1

引证文献1

二级引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部