期刊文献+

ICP-OES法和过硫酸铵分光光度法测定地表水中锰的方法比较 被引量:1

Comparison between ICP-OES and Ammonium Persulfate Spectrophotometry in the Determination of Manganese in Surface Water
下载PDF
导出
摘要 比较ICP-OES法与过硫酸铵分光光度法测定地表水中锰的差异性。结果显示,两种方法均线性良好,相关系数均为0.9999;有证的质控样测定结果在不确定度范围内。但当样品质量浓度小于0.05mg/L时,过硫酸铵分光光度法的精密度、回收率较差;ICP-OES法在高、中低浓度的测定精密度、准确度、回收率均较好。建议样品浓度低于0.05mg/L时,采用ICP-OES法测定样品。 Manganese in surface water was determined by ICP-OES and ammonium persulfate spectrophotometry methods.Linearity was good for manganese determined by both methods,with correlation coefficients 0.9999.RDSs of certified standard water samples were dropped in the ranges of uncertainty.However the precisions and recovery of ammonium persulfate spectrophotometry were not satisfied,when the concentrations of Mn were less than 0.05 mg/L.Meanwhile,the precisions,accuracy and recovery of ICP-OES were all good no matter whether the concentrations of Mn were low or high.It is suggested that when the concentrations of Mn are less than 0.05 mg/L,the samples should be determined by ICP-OES.
作者 王振华 王铮 李宁 WANG Zhenhua;WANG Zheng;LI Ning(Shanghai National Engineering Research Center of Urban Water Resources Co.,Ltd.,Shanghai 200082,China)
出处 《净水技术》 CAS 2021年第S01期24-26,31,共4页 Water Purification Technology
基金 上海市2020年度“科技创新行动计划”社会发展科技攻关项目"上海市深层地下水特征污染物分析与处理技术研究"(20DZ1204504)。
关键词 电感耦合等离子体发射法 过硫酸铵分光光度法 ICP-OES ammonium persulfate spectrophotometry manganese
  • 相关文献

参考文献3

二级参考文献22

共引文献7

同被引文献2

引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部