期刊文献+

多针孔心脏SPECT成像系统设计与性能评估 被引量:5

System Design and Performance Evaluation for Cardiac SPECT Imaging with Multi-pinhole Collimator
下载PDF
导出
摘要 单光子发射计算机断层成像技术(SPECT)是放射性核素心脏显像重要的成像手段之一。受传统SPECT的平行孔准直器的灵敏度和空间分辨率的影响,SPECT心脏成像的图像质量不佳。本文在临床SPECT系统的双探头基础上,通过设计高灵敏度、高空间分辨率的多针孔准直器,取代传统的平行孔准直器,实现高性能的心脏成像。通过解析计算对系统性能进行了评估,结果表明:中心灵敏度为0.038%,空间分辨率为1 cm,优于传统的平行孔SPECT。基于数值模拟的心肌成像实验,获得了高质量的心肌重建图像,证明了基于多针孔准直器心脏SPECT成像的设计可行性。 SPECT is one of the powerful tools for radionuclide cardiac imaging.However,the image quality of cardiac SPECT imaging usually is degraded by the poor sensitivity and spatial resolution of conventional SPECT parallel-hole collimator.Based on the two detectors from a clinical SPECT scanner,multi-pinhole collimator was designed,which was used to replace the conventional parallel-hole collimator,for the cardiac imaging with high performance.The system performance was evaluated with analytical calculation.The results show that the central sensitivity is 0.038%,the spatial resolution is 1 cm,and it is better than that of the conventional SPECT.The projection with myocardial phantom was simulated.The reconstruction image with high quality was obtained by using iterative reconstruction algorithm.All the results demonstrate the feasibility of our proposed design of the cardiac SPECT with multi-pinhole collimator.
作者 孙立风 吕振雷 侯岩松 刘伟 江年铭 刘辉 马天予 SUN Lifeng;LYU Zhenlei;HOU Yansong;LIU Wei;JIANG Nianming;LIU Hui;MA Tianyu(CNNC High Energy Equipment(Tianjin)Co.,Ltd.,Tianjin 300300,China;Department of Engineering Physics,Tsinghua University,Beijing 100084,China;Beijing Novel Medical Equipment Ltd,Beijing 102206,China)
出处 《原子能科学技术》 EI CAS CSCD 北大核心 2021年第S02期407-413,共7页 Atomic Energy Science and Technology
基金 国防科工局核能开发项目 中核集团“青年英才”项目 国家重点研发计划(2019YFF0302503)。
关键词 心脏SPECT 高灵敏度 高空间分辨率 cardiac SPECT high sensitivity high spatial resolution
  • 相关文献

参考文献1

二级参考文献2

共引文献1

同被引文献34

引证文献5

二级引证文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部