摘要
A universal locking model for single ion optical clocks was built based on a simple integrator and a double integrator.Different integrator algorithm parameters have been analyzed in both numerical simulations and experiments.The frequency variation measured by the comparison of two optical clocks coincides well with the simulation results for different second integrator parameters.According to the experimental results,the sensitivity of the servo error influenced by laser frequency drift with the addition of a double integrator was suppressed by a factor of 107.In a week-long comparison of optical clocks,the relative uncertainty of the servo error is determined to be 1.9×10^(-18),which is meaningful for the systematic uncertainty of the transportable single^(40)Ca^(+)ion optical clock entering the 10^(-18)level.
作者
Jin-Bo Yuan
Jian Cao
Kai-Feng Cui
Dao-Xin Liu
Yi Yuan
Si-Jia Chao
Hua-Lin Shu
Xue-Ren Huang
袁金波;曹健;崔凯枫;刘道信;袁易;晁思嘉;舒华林;黄学人(State Key Laboratory of Magnetic Resonance and Atomic and Molecular Physics,Innovation Academy for Precision Measurement Science and Technology,Chinese Academy of Sciences,Wuhan 430071,China;Key Laboratory of Atomic Frequency Standards,Innovation Academy for Precision Measurement Science and Technology,Chinese Academy of Sciences,Wuhan 430071,China;University of Chinese Academy of Sciences,Beijing 100049,China)
基金
the National Key Research and Development Program of China(Grant No.2017YFA0304404)
the National Natural Science Foundation of China(Grant No.11674357)。