摘要
目的:探讨噪声指数结合多模型自适应迭代重建(ASIR-V)在学龄前儿童上腹部低剂量CT扫描中的应用价值,探寻适合学龄前儿童上腹部CT的最佳扫描方案。方法:连续选取在我院行上腹部CT检查的学龄前儿童50例,按噪声指数(NI)值不同随机分为5组(NI值分别设为9、11、13、15、17),每组10例。CT检查采用GE Revolution CT机,管电压100 kV,管电流50~350 mA,所有图像均行ASIR-V(权重分别为0%、20%、40%、60%、80%、100%,共6个亚组)。采用单因素方差分析比较不同NI和ASIR-V权重条件下图像CT容积剂量指数(CTDIvol)、剂量长度乘积(DLP)、标准差(SD)、信噪比(SNR)、对比信噪比(CNR)等客观指标,采用Kruskal-Wallis H检验比较图像主观评分。结果:随着NI值的增大,图像的SD、CNR、CTDIvol、DLP等逐渐降低,NI=15、17组的滤波反投影(FBP)重建图像评分低于3分;随着ASIR-V权重的增大,图像的SD、CNR等值逐渐升高,权重为40%和60%的图像显示效果最佳。NI=15且ASIR-V为40%、60%图像的主观评分大于3分,CTDIvol和DLP较NI=9组分别下降了约55.5%和54.4%。结论:学龄前儿童上腹部CT平扫采用低管电压100 kV结合NI=15及ASIR-V40%的扫描方案,可以在明显降低辐射剂量的同时获取不影响临床诊断的图像。
Objective:To explore the application of noise index combined with adaptive statistical iterative reconstruction Veo(ASIR-V)in low-dose CT scanning of upper abdomen of pre-school children.Methods:Fifty children with upper abdominal CT scanning in our hospital were enrolled and randomly divided into 5 groups according to NI=9~17(interval 2)with 10 children in each group.The tube voltage was 100kV and the tube current was 50~350 mA.All images were reconstructed with ASIR-V(0%~100%,interva 20%).The CTDIvol,DLP,SD,SNR,CNR and other indicators were analyzed and compared using one way ANOVA under the different conditions of NI and ASIR-V.And the subjective scores were compared by Kruskal-Wallis H test.Results:With the increase of NI level,the SD,CNR,CTDIvoland DLP decreased gradually,and the scores of FBP reconstructed images in NI=15 and 17 groups were lower than 3 points.With the increase of ASIR-V,the SD and CNR values gradually increased,and the images with ASIR-V 40%~60%were the best.The image of NI=15 with 40%~60%of ASIR-V scores were higher than 3 points,in comparison with NI=9 group,CTDIvol and DLP decreased by 55.5%and 54.4%,respectively.Conclusions:The low tube voltage 100kV combined with NI=15 and ASIR-V 40%used in upper abdomen CT scan of preschool children can significantly reduce the radiation dose and obtain the images without affecting the clinical diagnosis.
作者
魏伟安
杨吉钱
熊胜
易婷
金科
WEI Wei-an;YANG Ji-qian;XIONG Sheng(Department of Radiology,Hunan Children's Hospital,Changsha 410007,China)
出处
《放射学实践》
CSCD
北大核心
2021年第9期1137-1141,共5页
Radiologic Practice
关键词
儿童
噪声指数
多模型迭代重建
体层摄影术
X线计算机
辐射剂量
图像质量
Child
Noise index
Adaptive statistical iterative reconstruction veo
Tomography,X-ray computer
Radiation dose
Image Quality