期刊文献+

Recent impact study of automatic search function of the pacemaker atrial hysteresis on atrial arrhythmias

下载PDF
导出
摘要 Objective:To explore the recent influences of pacemaker with automatic search function of atrial hysteresis on atrial arrhythmias,and to evaluate it clinical efficacy and safety.Methods:Indentify ADx DDD 5286 implanted dual chamber pacemaker for sick sinus syndrome in 43 cases.Automatic search of atrial lag was not opened with after pacemaker implantation,and the pacemaker settings were kept.Follow-up program after 3 months,DDD mode with automatic search of atrial lag was opened,and this mode was followed up for 6 months,comparing the atrial pacing percentage and DCG atrial tachyarrhythmias of pacemaker implantation to opening atrium lag mode.Results:Compared with the preoperative and operative 3 months later,dynamic electrocardiogram(DCG)24 h showed that the number of atrial premature beats(APB)and atrial tachycardia,atrial fibrillation(AF)array increased(p<.05);the cases of APB,atrial tachycardia and AF episodes were also increased(p<.05).Compared with the automatic search function in atrial hysteresis model opened with and not opened:atrial pacing percentage decreased[0.54(0.41,0.71)vs.0.82(0.65,0.93),p<.05];DCG 24 h showed that the number of APB,AF episodes was reduced(p<.05).Conclusions:Automation search function in atrial hysteresis model can obviously reduce the proportion of atrial pacing,reduce the occurrence of atrial arrhythmias;opened with automatic search function in atrial hysteresis model was safe and reliable.
出处 《Discussion of Clinical Cases》 2016年第4期14-17,共4页 临床病例讨论杂志(英文版)
  • 相关文献

参考文献2

二级参考文献15

  • 1时向民,王玉堂,单兆亮,李莉,阎俊瑾.犬心房不同部位起搏对不应期离散度的影响[J].第二军医大学学报,2004,25(11):1253-1254. 被引量:5
  • 2郭继鸿.生理性起搏及临床应用[J].心血管病学进展,2006,27(2):144-148. 被引量:25
  • 3朱冰船,刘启明,杨宇,祁述善,周胜华.触发心房颤动的房性期前收缩特征及抗心房颤动起搏器的疗效[J].临床心血管病杂志,2007,23(3):174-176. 被引量:6
  • 4宿燕岗,葛均波.生理性起搏的再认识[J].中国心脏起搏与心电生理杂志,2007,21(3):196-199. 被引量:36
  • 5Nitardy A, Langreck H, Dietz R, et al. Reduction of right ventricular pacing in patients with sinus node dysfunction through programming a long atrioventricular delay along with the DDIR mode. Clin Res Cardiol, 2009, 98:25-32.
  • 6Mitro P, Kotianova A, Bodnar J, et al. Quality of life and psychological well-being in patients with various pacing modes. Bratisl Lek Listy,2008, 109:260-266.
  • 7Mont L,Ruiz-Granell R,Martinez JG,et al. Impact of anti-tachycardia pacing on atrial fibrillation burden when added on top of preventive pacing algorithms: results of the prevention or termination (POT) trial. Europace, 2008,10:28-34.
  • 8Van Eck Jw,Van Hemel Nm,Van den Bos A,et al. Predictors of improved quality of life 1 year after pacemaker implantation.Am Heart,2008, 156:491-497.
  • 9Deharo JC, Djiane P.Pacemaker longevity. Replacement of the device. Ann Cardiol Angeiol,2005,54:26-31.
  • 10Sweeney MO, Bank AJ, Nsah E, et al.Minimizing ventricular pacing to reduce atrial fibrillation in sinus-node disease. N Engl J Meal, 2007,6:1000-1008.