期刊文献+

RF power design optimization in MRI system 被引量:1

下载PDF
导出
摘要 Magnetic resonance image quality and patient safety have been the focus of engineering and research ever since the invention of equipment in the early 1970s.In high field(or ultrahigh field)MRI systems,the emerging techniques induced by B1 field challenges have promoted various potential solutions.This paper describes the relationship between RF power and B1þfield performance,and the overall requirements considered in RF subsystem design.The design of the RF in the MR system is systematically summarized,including the entire transmission chain,sequence algorithm and RF pulse design,and the probabilities for improvement and optimization in the system design are indicated.At the same time,the radio frequency related issues of the human whole-body 7 T MR and animal MR systems are discussed,especially the promising future showed by the technologies such as radio frequency parallel transmission technology in the ultrahigh field.
出处 《Magnetic Resonance Letters》 2021年第1期89-98,共10页 磁共振快报(英文)
基金 The work is supported by Key-Area Research and Development Program of Guangdong Province(grant no.2018B030333001).
  • 相关文献

同被引文献3

引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部