期刊文献+

实验室用纯水机校准方法探索

Research of Calibration Method for Laboratory Water Purification Machine
下载PDF
导出
摘要 实验室用纯水机是采用预处理、反渗透技术、超纯化处理以及后级处理等步骤,将水中的导电介质几乎完全去除的水处理设备,其因为多数采用一体化设计,且电导率/电阻率探头不易拆卸,因此无法用常规方法对电导率/电阻率值进行校准。本文介绍了一种比对实验方法,对实验室用纯水机电导率/电阻率值进行计量,并采用称量法,对仪器流量进行计量,为广大科技工作者提供了参考。 Laboratory water purification machine is a kind of water treatment equipment,which uses pretreatment,reverse osmosis technology,super-purification treatment and post-treatment to remove the conductive medium in water almost completely,because most of them are of integrated design,the conductivity/resistivity probe is not easy to disassemble and can not be calibrated by conventional methods,in this paper,a method of comparison ex-periment is introduced,which is used to measure the conductivity/resistivity of pure water machine in laboratory and to measure the flow of instrument by weighing method,the experimental method provides the reference for the general scientific and technological workers.
作者 何枫 张平 刘东晖 吴聪 HE Feng;ZHANG Ping;LIU Donghui;WU Cong
出处 《计量与测试技术》 2021年第9期61-64,共4页 Metrology & Measurement Technique
关键词 实验室用纯水机 电导率 电阻率 比对试验 示值误差 laboratory water purification machine conductivity resistivity comparison test indication error