摘要
目的:探究分析酶标法测定人血清中HIV抗体的不确定度。方法:2019年8月~2019年12月根据JJF1059-1999《测量不确定度评定与表示》,采用酶标法测定人血清中HIV抗体,通过HIV抗体检查结果,明确其不确定度,并进行详细分析。结果:资料显示,阴性B=0.28,阴性吸光度值=0.078,cutoff值=0.278。通过cutoff值可将说明书中的批内精密性原始值重新还原成吸光度值。结合微量加样器校准证书可得出,75μ1的相对误差是0.15%,可知不确定度=0.234μ1;结合75μ1微量加样器在试验期间核查的相关数据,可知不确定度=0.341μ1;结合检验规程的规定,每批次检验过程中样品的加样时间均小于30min,可知不确定度=8.66min;结合酶标仪鉴定证书,可知不确定度=0.01。最终可得,样品的吸光度值和cutoff值比值B=4.332,不确定度U=1.2,k=2。结论:通过酶标法测定人血清中HIV抗体的不确定度,可减少假阴性出现几率,可辨别检测质量的影响因素,有利于实验室检测质量的提高。
Objective:Exploring and analyzing the uncertainty of the determination of HIV antibody in human serum by enzyme-labeled method.Methods:From August 2019 to December 2019,according to JJF1059-1999"Evaluation and Expression of Uncertainty in Measurement",the enzyme-labeled method was used to determine the HIV antibody in human serum,and the result of HIV antibody test was used to clarify the uncertainty and carry out Detailed analysis.Results:The data showed that negative B=0.28,negative absorbance value=0.078,cutoff value=0.278.The cutoff value can restore the original value of the intra-batch precision in the manual to the absorbance value.Combining with the micro-sampler calibration certificate,it can be concluded that the relative error of 75μ1 is 0.15%,and the uncertainty is 0.234μ1;combined with the relevant data checked by the 75μ1 micro-sampler during the test,the uncertainty=0.341μ1;in accordance with the inspection regulations,the sample addition time for each batch of samples during the inspection process is less than 30 minutes,and the uncertainty=8.66 minutes;combined with the identification certificate of the microplate reader,the uncertainty=0.01.Finally,it can be obtained that the ratio of the absorbance value and the cutoff value of the sample is B=4.332,the uncertainty U=1.2,and k=2.Conclusion:The determination of the uncertainty of HIV antibodies in human serum by the enzyme-labeled method can reduce the probability of false negatives and identify the factors affecting the quality of the test,which is beneficial to the improvement of the quality of laboratory tests.
作者
石冬月
SHI Dong-yue(Center for Disease Control and Prevention,Wanghua District,Fushun City,Liaoning Province,Liaoning Fushun 113001)
出处
《中国医疗器械信息》
2021年第18期54-55,共2页
China Medical Device Information
关键词
酶标法
HIV抗体
不确定度
enzyme labeling method
HIV antibodies
the uncertainty