缝管法增强火焰原子吸收分析灵敏度的研究——测定血清中铜和锌 被引量:7

The Enhance of the Sensitivity of Conventional Flame Atomic-Absorption Spectrometry by Slotted Stainless Steel--Determination of Copper and Zinc in Serum
下载PDF
导出
摘要 火焰原子吸收光谱法由于基体干扰少、操作简便、设备廉价得到了广泛应用,但同时又因为灵敏度不够高其应用受到某些限制,非火焰的石墨炉原子吸收分析法,灵敏度较火焰法有很大提高.但由于有较严重的基体效应、设备复杂、价格昂贵。 A slotted stainless steel tube when used in air-acetylene flame increses the residence time of atoms in the light path of an atomic-absorption spectrometer,thus increasing the analytical sensitivity.With this technique copper and zinc were determined in serum diluted 1:20.Rapid accurate analyses can be achieved by using continuous aspiration.Use of aqueous standards resulted in recovery for copper and zinc of 92-110.9%.Coefficient of variation of 2.7% and 3.6%respectively.
作者 孙书菊 邴贵德 郑衍生 郭毅 Sun Shuju;Bing Guide;Zheng Yansheng;Guo Yi(Department of Chemistry,Jilin University,Changchun)
机构地区 吉林大学化学系
出处 《高等学校化学学报》 SCIE EI CAS 1987年第11期978-980,共3页 Chemical Journal of Chinese Universities