东芝Aquilion one CT维修案例分析与解决

Toshiba Aquilion one CT Maintenance Case Analysis and Solution
下载PDF
导出
摘要 本文分析了东芝Aquilion one CT的三例故障,分别为出现曝光旋转时,机架噪声大;重建柜内QFPGA功能板件损坏,冠脉重建时出现报错;球管损坏,设备曝光失败,报“Over mA”故障。采用了逐步排除法、仔细观察法、软件数据分析法等检修方法,通过分别对机架轴承添加润滑油、更换QFPGA板件、更换同型号球管进行参数调试校准后,故障得到解决。这些检修方法的运用及相应解决问题的方案,给同行提供了参考与借鉴,对于提高维修效率,为医院节省维修成本和保证医院相关科室工作正常开展,发挥了重要作用。 This article analyzed three failures of Toshiba Aquilion one CT.They were the loud noise of the rack when the exposure rotation occured;the QFPGA function board in the reconstruction cabinet was damaged,and an error occured during coronary reconstruction;the tube was damaged and the equipment failed to be exposed,and reported“Over mA”.After adopting maintenance methods such as step-by-step elimination method,careful observation method,and software data analysis method,the faults were solved by adding lubricating oil to the rack bearing,replacing the QFPGA board,and replacing the same type of tube for parameter debugging and calibration.The application of these maintenance methods and corresponding problem-solving programs have provided reference and reference for colleagues,and play an important role in improving maintenance efficiency,saving maintenance costs for the hospital,and ensuring the normal operation of related departments in the hospital.
作者 吴观友 WU Guanyou(Department of Equipment,The First Affiliated Hospital of Jinan University,Guangzhou Guangdong 510630,China)
出处 《中国医疗设备》 2021年第11期170-172,176,共4页 China Medical Devices
关键词 东芝Aquilion one CT 机架轴承 QFPGA板件 曝光球管 Toshiba Aquilion one CT frame bearing QFPGA board exposure tube
  • 相关文献

参考文献11

二级参考文献60

  • 1常少卿,孔江涛.岛津XUD150L-F X线机高压故障二例[J].现代医用影像学,2011,20(4):275-276. 被引量:2
  • 2张永顺,杜昱平.FSK302-Ⅰ型程控X线机采样电路故障二则[J].临床放射学杂志,2006,25(8):715-715. 被引量:4
  • 3章祖华,吴川军,吴明兰.X线机主要参数的检测、评价及应用[J].医疗卫生装备,2007,28(7):64-65. 被引量:3
  • 4Schoepf UJ, Costello P. Multidetector-row CT imaging of pulmonary embolism. Semin Roentgenol, 2003,38(2) : 106-114.
  • 5Schmidt B. Dual source CT technology//Seidensticker PR. Dual source CT imaging. Heidelberg, Germany: Springer Berlin, 2008 : 19-33.
  • 6Greess H, Wolf H, Baum U, et al. Dose reduction in computed tomography by attenuation-based on-line modulation of the tube current : evaluation of six anatomical regions. Eur Radiol, 2000,10 (2) :391-394.
  • 7Millner MR, McDavid WD, WaggenerRG, et al. Extraction of information from CT scans at different energies. Med Phys, 1979,6 (1):70-71.
  • 8Jakobs TF, Becker CR, Ohnesorge B, et al. Multislice helical CT of the heart with retrospective ECG gating: reduction of radiation exposure by ECG-controlled tube current modulation. Eur Radiol, 2002,12(5) :1081-1086.
  • 9McCollough CH, Bruesewitz MR, Kofler JM Jr, et al. CT dose reduction and dose management tools: overview of available options. Radiographics, 2006,26(2):503-512.
  • 10Johnson TR, Krauss B, Sedlmair M,et al. Material differentiation by dual energy CT: initial experience. Eur Radio, 2007, 17 (6):1510-1517.

共引文献48

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部