摘要
目的:比较VX120多功能眼检查仪与TopconKR800全自动验光仪的客观验光结果的一致性。方法:收集2019年11月至2019年12月在我院就诊拟行角膜屈光手术的患者24例(24只眼),所有患者均取右眼。采用TopconKR800全自动验光仪和VX120多功能眼检查仪分别对24例患者进行客观验光检查,分别得到球镜光度(S)、柱镜光度(C)、等效球镜光度(M)、水平和垂直方向的柱镜光度(J0)及45°柱镜光度(J45)。采用配对T检验对两种设备的测量值进行组间比较,并采用Bland-Altman 95%一致性界限法衡量其一致性。结果:VX120多功能眼检查仪在明视模式下所测S为(−3.90±3.06)D;M为(−4.41±3.24)D;J0为(0.34±0.58)D;J45为(−0.04±0.26)D;TopconKR800全自动验光仪所测S为(−3.77±2.88)D;M为(−4.15±3.07)D;J0为(0.25±0.47)D;J45为(−0.01±0.18)D。VX120多功能眼检查仪在明视模式与TopconKR800全自动验光仪所测得的S、M、J45差异无统计学意义,而J0两者之间差异有统计学意义(t=−2.24,P=0.04)。Bland-Altman 95%一致性分析显示,两种仪器参数数据点多分布在95%可信区间内,一致性较好。结论:VX120多功能眼检查仪明视模式下与TopconKR800全自动验光仪的客观验光结果一致性较好,在屈光度检查上可以相互替代。
Objective:The aim of this study was to compare the consistency of objective optometry results between VX120 multi-diagnostic unit and TopconKR800 automatic optometer.Methods:A total of 24 patients(24 right eyes)admitted to our hospital from November 2019 to December 2019 for corneal refractive surgery were collected.Optometry was performed on 24 patients with TopconKR800 automatic optometer and VX120 multi-diagnostic unit,and got the results following:spherical mirror photometry(S),cylindrical lens luminosity(C),mean spherical equivalent(M),horizontal and vertical cylindrical lens luminosity(J0)and 45°cylindrical lens luminosity(J45).Compared T-test analysis was used to compare the parameters of the two devices,and the Bland-Altman 95%limits of agreement were performed.Results:The S measured by the VX120 multi-diagnostic unit with bright mode was(−3.90±3.06)D;M was(−4.41±3.24)D;J0 was(0.34±0.58)D;J45 was(−0.04±0.26)D.The S measured by TopconKR800 automatic optometer was(−3.77±2.88)D;M was(−4.15±3.07)D;J0 was(0.25±0.47)D;J45 was(−0.01±0.18)D.There was no statistically significant difference in S,M,and J45 measured by VX120 multi-diagnostic unit and TopconKR800 automatic optometer in bright mode,while there was statistically significant difference in J0 between VX120 multi-diagnostic unit and TopconKR800 automatic optometer(t=−2.24,P=0.04).Bland-Altman 95%consistency analysis showed that the data points of the two instruments were mostly distributed within the 95%confidence interval,indicating that the parameters measured by these two devices were in accord with each other.Conclusion:The objective optometry results of VX120 multi-diagnostic unit and TopconKR800 automatic optometer were consistent and can be substituted for each other in the refraction examination.
作者
林慧妮
刘秋霞
孙丽霞
LIN Huini;LIU Qiuxia;SUN Lixia(Joint Shantou International Eye Center,Shantou University and The Chinese University of Hong Kong,Shantou Guangdong 515041,China)
出处
《眼科学报》
CAS
2021年第10期787-791,共5页
Eye Science