期刊文献+

基于心冲击信号的心率检测方法

Heart rate detection based on ballistocardiogram signals
下载PDF
导出
摘要 针对现有基于心冲击描记图进行心率检测存在的抗干扰性不足问题,提出一种鲁棒的心率检测模型,首先对原始信号中受干扰的片段进行补偿重构,然后将J峰作为特征波形,通过基于J峰频段的滤波与差分增强处理放大特征,并设计改进的峰值提取算法定位心跳点,进而得到心率估计值。实验结果表明,所用模型对于不同情形下的心冲击描记图信号均表现出良好的适应性,逐拍检测平均覆盖率为98.3%,平均准确率为98.4%。心率输出结果与金标准一致性高,误差范围控制在±5%以内,且运算实时性好,可以为临床或家庭保健提供良好的心率检测方案。 In view of the lack of anti-interference in existing heart rate detection based on ballistocardiograms,a robust heart rate detection model is proposed.After compensating the interfered fragments in the original signal,J-peak is taken as feature waveform,and J-wave features are amplified based on the filtering of J-peak frequency band and enhanced differential processing.Finally,an improved peak extraction algorithm is designed to locate the heartbeat and then obtain the estimation of heart rate.The experimental results show that the model used in the study has good adaptability to ballistocardiogram signals in different situations.The average coverage rate of beat by beat detection is 98.3%,and the average accuracy reached 98.4%.The heart rate output is in high consistency with gold standard,with an error controlled within±5%.Moreover,the real-time operation is satisfactory.The proposed method can provide an appropriate heart rate detection scheme for clinical or family health care.
作者 郑小涵 朱岩 杨越琪 李晓欧 周志勇 ZHENG Xiaohan;ZHU Yan;YANG Yueqi;LI Xiaoou;ZHOU Zhiyong(School of Medical Instrument and Food Engineering,University of Shanghai for Science and Technology,Shanghai 200093,China;College of Medical Instruments,Shanghai University of Medicine&Health Sciences,Shanghai 201318,China;School of Design and Art,Shanghai Dianji University,Shanghai 200240,China)
出处 《中国医学物理学杂志》 CSCD 2021年第11期1405-1411,共7页 Chinese Journal of Medical Physics
基金 上海市科委科技创新行动计划(19441914900)。
关键词 心冲击描记图 心率检测 非接触式 峰值检测 ballistocardiogram heart rate detection non-contact peak detection
  • 相关文献

参考文献9

二级参考文献91

共引文献64

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部