期刊文献+

全自动微生物分析仪准确性的影响因素及对策分析 被引量:1

Influencing Factors and Countermeasures of Accuracy of an Automatic Microbial Analyzer
下载PDF
导出
摘要 目的探讨VITEK 2 Compact全自动微生物分析仪准确性的影响因素及对策。方法对三例血培养的错误鉴定结果进行综合原因分析,从硬件设施、试剂耗材、操作程序等方面寻找鉴定错误的原因,对试剂耗材用标准菌株进行内质控程序。结果发现竹杆棉拭子制备上机鉴定的菌悬液会对VITEK 2 Compact的鉴定结果造成严重影响,菌株不纯、鉴定卡选择错误、菌悬液制备不当、超出鉴定范围等也会发生不同程度影响。结论使用全自动微生物分析仪时要严格按照操作要求进行,综合分析结果,对可疑结果应主动复核。 Objective To explore the influencing factors and Countermeasures of the accuracy of VITEK 2 Compact automatic microbiological analyzer.Methods The causes of misidentification of three blood cultures were analyzed comprehensively.The reasons of misidentification were found from hardware facilities,reagent consumables,operation procedures and so on.The internal quality control program of standard strains for reagent consumables was carried out.Results It was found that the bacterial suspension prepared by bamboo cotton swab could seriously affect the identification results of VITEK 2 Compact.The impurity of strains,wrong card selection,improper preparation of bacterial suspension and beyond the scope of identification would also have different effects.Conclusion When using automatic microbiological analyzer,the operation requirements should be strictly followed,and the suspicious results should be checked actively by synthesizing the analysis results.
作者 张迎华 张保荣 ZHANG Yinghua;ZHANG Baorong(Department of Laboratory Medicine,Central Hospital of Datun Coal and Electricity(Group)Co.Ltd.,Xuzhou Jiangsu 221611,China;Department of Laboratory Medicine,Suqian First Hospital,Suqian Jiangsu 223800,China)
出处 《中国继续医学教育》 2021年第33期111-114,共4页 China Continuing Medical Education
关键词 微生物鉴定 药敏 影响因素 室内质控 棉拭子 全自动微生物分析仪 microbial identification drug sensitivity influencing factors indoor quality control cotton swab control automatic microbial analyzer
  • 相关文献

参考文献15

二级参考文献111

共引文献96

同被引文献13

引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部