基于美国放射学会标准的MRI放射治疗模拟定位系统长期稳定性检测

Detection on the long term stability of simulation positioning system of MRI radiotherapy based on ACR standard
下载PDF
导出
摘要 目的:基于美国放射学会(ACR)模体检测MRI定位系统的稳定性和图像质量,并将该方法纳入质量控制流程,以保证临床扫描数据的可靠性。方法:收集2019-2020年(每月1次,共12次)内在德国西门子3.0T Skyra大孔径MRI定位机上的T_(1)、T_(2)序列扫描ACR Large Phantom模体图像数据。根据ACR模体指南,分析比较ACR Large Phantom模体图像数据与MRI定位系统扫描图像的几何精度(误差±2 mm)、高对比度空间分辨率(≥1 mm)、图像均匀性(≥82%)、信号伪影率(≤0.025)和低对比度分辨率[可观测总轮辐数T_(1)加权成像(T_(1)WI)≥35,T_(2)WI≥31]。结果:矢状位、T_(1)WI和T_(2)WI图像的几何精度差异均<±1.5 mm;T_(1)WI和T_(2)WI图像的高对比度空间分辨率为1.0 mm,图像均匀度百分比(PIU)平均值分别为88.86%和85.00%,平均信号伪影分别为0.0014和0.0035,测量低对比度分辨率的可观测总轮辐数均值分别为39和37。各评价指标均满足临床图像质量需求和ACR模体指南处置限值。结论:基于ACR模体掌握MRI图像质量检测方法,并定期进行图像质量检测,是保证MRI系统稳定和图像质量的重要质量控制手段,可用于长期检测MRI系统和成像质量,不断提高临床诊疗水平。 Objective:To detect the stability and image quality of magnetic resonance imaging(MRI)positioning system based on the phantom of American College of Radiology(ACR),and bring this method into the process of quality control so as to ensure the reliability of clinical scanning data.Methods:The image data of ACR Large Phantom with T_(1) and T_(2) sequence scan on Siemens 3.0T Skyra large aperture MRI positioning machine were collected during 2019 and 2020(once a month,total of 12 times).According to the guideline of ACR phantom to analyze and compare the geometric accuracy(error±2 mm),high-contrast spatial resolution(≥1 mm),image homogeneity(≥82%),percent-signal ghosting(≤0.025)and low-contrast resolution(the number of observable total spokes T_(1)WI≥35,T_(2)WI≥31).Results:The differences of geometric accuracies among sagittal view and the images of T_(1)WI and T_(2)WI were less than±1.5 mm.The high-contrast spatial resolutions of T_(1)WI and T_(2)WI images were 1.0 mm respectively,and the means of the percentage of image uniformity(PIU)of them were 88.86% and 85.00% respectively,and the average presents-signal ghosting of them were 0.0014 and 0.0035 respectively.The average number of observable total spokes were 39 and 37 respectively in detecting low-contrast resolution.The results indicated that each evaluation indicator could meet the requirement of clinical image quality and the limiting value of ACR phantom guideline.Conclusion:The grasped detection method of MRI image quality on the basis of ACR phantom and the regular detection of image quality are an important measure of quality control in ensuring the stability of MRI system and image quality,which can be used in long term detection of MRI system and imaging quality so as to improve continually the clinical level of diagnosis and treatment.
作者 李莎 吴昊 刘长斌 胡俏俏 刘巍巍 张艺宝 LI Sha;WU Hao;LIU Chang-bin(Department of Radiotherapy,Beijing Cancer Hospital of Peking University/Beijing Institute for Cancer Research,Key Laboratory of Carcinogenesis and Translational Research,Ministry of Education,Beijing 100142,China;不详)
出处 《中国医学装备》 2022年第1期49-53,共5页 China Medical Equipment
基金 国家自然科学基金(11505012,11905150)“低剂量锥束CT扫描参数个体化优化算法”“用于图像引导放射治疗的双能锥束CT的个体化成像方案研究” 国家重点研发项目(2019YFF01014405)“质子治疗剂量验证与治疗计划系统的研制” 首都卫生发展科研专项(首发2018-4-1027)“基于患者实际体位修正滑环加速器兆伏锥束CT成像剂量” 北京市自然科学基金(7172048)“基于先验知识的智能放疗计划设计建模研究” 教育部科技发展中心产学研创新基金(2018A01019)“基于京津冀地区食管癌放疗大数据模型和递进算法的精确治疗方案交互优化” 北京市属医院科研培育计划(PX2019042)“基于人工智能的调强放疗计划优化参数自动调整方法研究” 北京大学医学部教育教学研究立项(2020YB34)“人工智能辅助的肿瘤放疗方案设计教学方法”。
关键词 磁共振成像(MRI) 美国放射学会(ACR) 模体 质量控制 图像质量 稳定性 Magnetic resonance imaging(MRI) American College of Radiology(ACR) Phantom Quality control Image quality Stability