微波消解-碰撞池-电感耦合等离子体质谱(ICP-MS)法测定芝麻中痕量的锗 被引量:3

Determination of Trace Germanium in Sesame by Inductively Coupled Plasma-Mass Spectrometry-Collision Cell Technology with Microwave Digestion
下载PDF
导出
摘要 建立了微波消解-碰撞池-电感耦合等离子体质谱法测定芝麻中痕量锗元素的测定方法。在线引入内标元素铑(Rh),同时消解液中加入正戊醇(3%)作为增敏剂。结果显示,正戊醇(3%)可使^(74)Ge的上机检测信号强度提高2.85倍,^(74)Ge校正曲线线性相关系数为1.000,检出限为0.056μg/kg,加标回收率为92.0%~106%,相对标准偏差(RSD)为2.6%~4.3%。采用建立的方法测定7种国家标准物质,检测结果均在认定值范围内,RSD为2.5%~8.8%。方法灵敏度高、准确,可实现批量检测,适用于芝麻中痕量锗的检测。 In order to accurately and efficiently detect trace germanium(Ge)in sesame,a method of microwave digestion-collision cell-inductively coupled plasma mass spectrometry was established.The internal standard element of rhodium(Rh)was introduced online,and addition of 3%1-pentanol to the digestion solution were used to enhance the sensitivity.The results indicated:3%1-pentanol increased the signal intensity of ^(74)Ge by 2.85 times.The correlation coefficient of germanium correction curve was 1.000.The limit of detection of the method was 0.056μg/kg,and the spiked recovery rates ranged from 92.0%to 106%with the relative standard deviation(RSD)of 2.6%—4.3%.The certified reference material of germanium was in the range of certified value.This method has high sensitivity,high accuracy and good precision,and is suitable for the determination of germanium element content in sesame.
作者 巩佳第 段晓婷 章路 孙玉梅 郭利攀 余程凤 GONG Jiadi;DUAN Xiaoting;ZHANG Lu;SUN Yumei;GUO Lipan;YU Chengfeng(Greentown Agrotech Testing Technology Co.,Ltd.,Hangzhou,Zhejiang 310052,China)
出处 《中国无机分析化学》 CAS 北大核心 2022年第1期133-138,共6页 Chinese Journal of Inorganic Analytical Chemistry
基金 国家重点研发计划资助项目(2018YFC1602800)。
关键词 碰撞池 电感耦合等离子体质谱法 芝麻 正戊醇 collision cell technology inductively coupled plasma-mass spectrometry sesame germanium 1-pentanol
  • 相关文献

参考文献20

二级参考文献227

共引文献151

同被引文献46

引证文献3

二级引证文献13

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部