噪声指数在CT定位肺部穿刺活检中的应用 被引量:2

The application of noise index in CT localization lung puncture biopsy
下载PDF
导出
摘要 目的:探讨螺旋CT噪声指数技术(noise index,NI)在肺部CT引导下穿刺定位的应用价值。方法:选取30例进行CT引导下穿刺活检的患者,首次扫描采用胸部剂量(60 kV、100 mAs,NI20)扫描滤波反投影重建,后续监测穿刺针插入肿瘤过程扫描依次采用较高噪声(NI40)和高噪声指数(NI60)。记录每次扫描的容积CT剂量指数,剂量长度乘积,并计算有效剂量。结果:三种噪声指数剂量比较辐射剂量分别降低了47%与77%。3组的图像质量均能满足穿刺活检检查监测穿刺针的诊断要求。结论:将噪声指数技术应用到低剂量CT引导下穿刺活检定位中,可以大幅降低辐射剂量,值得临床推广。 Objective:To explore the value of spiral CT noise index technique in lung CT guided puncture localization.Methods:Thirty patients were selected for CT-guided puncture biopsy.The first scan was performed with chest dose(60 kV,100 mAs,NI20)scan filtering and back projection reconstruction,followed by follow-up monitoring of puncture needle insertion process and followed by higher noise(NI40)and highest noise index(NI60).The volumetric CT dose index,dose length product and effective dose were recorded for each scan.Results:The radiation doses of the three noise index doses were reduced by 47%and 77%respectively.The image quality of the three groups can meet the diagnostic requirements of puncture biopsy monitoring needle.Conclusion:The application of noise index technique to low dose CT guided puncture biopsy localization can significantly reduce the radiation dose,which is worthy of clinical promotion.
作者 刘晨鹏 潘振宇 闫玉昌 李坤 LIU Chenpeng;PAN Zhenyu;YAN Yuchang;LI Kun(Department of Radiology,Beijing Chaoyang Hospital,Capital Medical University,Beijing 100040,China)
出处 《现代肿瘤医学》 CAS 北大核心 2022年第1期117-120,共4页 Journal of Modern Oncology
关键词 噪声指数 穿刺定位 肺肿瘤 辐射剂量 noise index puncture positioning lung cancer radiation dose
  • 相关文献

参考文献13

二级参考文献75

  • 1曲顺福.超低剂量CT在肋骨骨折诊断中的可行性分析[J].世界临床医学,2017,11(3):227-227. 被引量:1
  • 2倪颖梦,时国朝,万欢英,陈克敏,吴达明.CT引导下经皮肺穿刺的安全性及其影响因素[J].中国呼吸与危重监护杂志,2011,10(2):162-167. 被引量:32
  • 3王焕玲,李太生,王爱霞,盛瑞媛,范洪伟,刘正印,滑立伟,秦树林,邓国华.艾滋病合并肺孢子菌肺炎22例临床分析[J].中华内科杂志,2005,44(9):652-655. 被引量:34
  • 4GBZ165-2012,X射线计算机断层摄影放射防护要求[s].
  • 5钟涛,于红光,王勇,杨思福,王晓轩.CT引导下经皮肺穿刺活检术后气胸发生率的相关因素分析[J].中华放射学杂志,2007,41(11):1232-1236. 被引量:56
  • 6Goo H W.CT radiation dose optimization and estimation:an update for radiologists[J].Korean J Radiol,2012,13(1):1-11.
  • 7Brenner D J,Hall E J.Computed tomography—an increasing source of radiation exposure[J].N Eng J Med,2007,357:2 277-2 284.
  • 8Kalra M K,Woisetschlager M,Dahlstrom N,et al.Radition dose reduction with sinogram affirmed iterative reconstruction technique for abdominal computed tomography[J].Comput Assist Tomogr,2012,336:339.
  • 9PAN Xiao-chuan,Sidky E Y,Vannier M.Why do commercial CT scanners still employ traditional,filtered back-projection for image reconstruction?[J].Inverse Probl,2009,25(12):1 230 009-1 230 059.
  • 10HU X H,DING X F,WU R Z,et a1.Radiation dose of non-enhanced chest CT can be reduced 40%by using iterafive reconstruction in image space[J].Clin Kadiol,2011,66:1 023-1 029.

共引文献64

同被引文献23

引证文献2

二级引证文献17

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部