期刊文献+

迈瑞体外除颤器的质量检测

Quality Inspection of Mindray Automated External Defibrillator
下载PDF
导出
摘要 目的探讨迈瑞体外除颤器(AED)的质量检测方法。方法按照JJF 1149-2014《心脏除颤器校准规范》、JJG 760-2003《心电监护仪检定规程》、GB 9706.8-2009《医用电气设备第2-4部分:心脏除颤器安全专用要求》的规定,结合使用说明书,使用福禄克(FLUKE)专业质量检测设备对院内的3台迈瑞半自动AED进行检测。结果根据校准规范及机型,制作了AED专用质量检测原始记录表,3台迈瑞半自动AED检测合格率均为100%。结论AED的质量检测关系重大,应加强对其的日常管理工作,保障设备随时处于待用状态。 Objective The quality inspection method of Mindray automated external defibrillator(AED)was explored.Methods According to JJF 1149-2014 Calibration Specification for Cardiac Defibrillators,JJG 760-2003 Verification Regulation for Electro Cardiac Monitor,and GB 9706.8-2009 Medical electrical equipment-Part 2-4:Particular requirements for the safety of cardiac defibrillators,and in combination with the instruction manual of Mindray AED,three Mindray AEDs in the hospital were tested by the FLUKE professional quality inspection equipment.Results According to the calibration specifications and models,the original quality inspection records for AED were made,and the qualified rate of the three inspected Mindray AEDs(semi-automatic)was 100%.Conclusions The quality inspection of AEDs is very important.Therefore,the daily management of defibrillators should be strengthened to ensure their standby status.
作者 刘富桃 蔡本忠 周伟 朱昌盛 Liu Futao;Cai Benzhong;Zhou Wei;Zhu Changsheng(Department of Medical Equipment,The Yancheng School of Clinical Medicine of Nanjing Medical University Yancheng Third People's Hospital,Yancheng Jiangsu 224001,China)
出处 《医疗装备》 2022年第1期48-50,共3页 Medical Equipment
关键词 自动体外除颤器 质量检测 电击心率识别 心电仿真信号 Automated external defibrillator Quality inspection Electric shock heart rate recognition Electrocardiogram simulation signal
  • 相关文献

参考文献4

二级参考文献28

共引文献13

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部