摘要
采用顶空-气相色谱法测定固定污染源废气中甲胺、二甲胺、乙胺、二乙胺和三乙胺,通过优化相关测定条件,使5种挥发性脂肪胺在0.005 mg/L~25.0 mg/L范围内线性良好,方法检出限为0.004 mg/m^(3)~0.4 mg/m^(3)(以采集20 L空气样品计)。5种挥发性脂肪胺的加标回收率为91.0%~113%,6次测定结果的RSD≤7.4%。
Methylamine,dimethylamine,ethylamine,diethylamine and triethylamine in waste gas from stationary pollution source were determined simultaneously by headspace-gas chromatography.By optimizing the relevant determination conditions,the linearity of five volatile fatty amines was good in the concentration range of 0.005 mg/L~25.0 mg/L,the method detection limits were between 0.004 mg/m^(3) and 0.4 mg/m^(3)(calculated by 20 L sampling volume).The recoveries of the five volatile fatty amines ranged from 91.0%to 113%,and the RSDs of six measurements were less than 7.4%.
作者
张凤菊
曹方方
岳太星
石锡峰
李红莉
侯聪
ZHANG Feng-ju;CAO Fang-fang;YUE Tai-xing;SHI Xi-feng;LI Hong-li;HOU Cong(Shandong Provincial Eco-environment Monitoring Centre,Jinan,Shandong 250101,China;Shandong Normal University,Jinan,Shandong 250014,China)
出处
《环境监测管理与技术》
CSCD
2022年第1期52-55,共4页
The Administration and Technique of Environmental Monitoring
基金
国家自然科学基金青年基金资助项目(21507074)。
关键词
挥发性脂肪胺
顶空-气相色谱法
固定污染源
废气监测
Volatile fatty amines
Headspace-gas chromatography
Stationary source
Waste gas monitoring