感兴趣体积IVS技术在腹部肿瘤配准范围的研究 被引量:1

Influence of IVS Different Registration Ranges in Abdomen Image Guidance
下载PDF
导出
摘要 目的采用苏州雷泰公司的电子射野影像系统(Image Verification System,IVS),探讨三种不同的感兴趣体积(Volume of Interest,VOI)配准范围在腹部肿瘤摆位验证中配准精度的影响。方法分别获取盆腔仿真模体的定位图像和20组已知位置偏差的IVS图像进行配准,选取三种不同的VOI范围,得到三维平移方向的摆位误差,进行对比分析。结果选取配准范围VOI为35 cm×35cm×35 cm摆位误差最小。VOI为35 cm×35 cm×35 cm与VOI为25 cm×25 cm×25 cm和10 cm×10 cm×10 cm相比,x、y、z三个方向的摆位误差均具有明显的统计学差异(P<0.05)。结论对于盆腔部肿瘤而言,图像配准范围VOI建议选取35 cm×35 cm×35 cm配准结果的误差最小。 Objective The effects of three different volume of interest(VOI)registration ranges on the accuracy of abdominal tumor placement verification were investigated by using image verification system(IVS)developed by Suzhou Reitel.Methods The positioning images of anthropomorphic pelvic phantom and 20 groups of IVS images with known position deviations were obtained for registration.Three different VOI ranges were selected to obtain the positioning errors of three-dimensional translation direction for comparative analyzed.Results The VOI of 35 cm×35 cm×35 cm was selected as the minimum positioning error.Compared with VOI of 25 cm×25 cm×25 cm×25 cm and 10 cm×10 cm×10 cm,the positioning errors in x,y and z directions had significant statistical differences(P<0.05).Conclusion For pelvic tumors,VOI of image registration range 35 cm×35 cm×35 cm is recommended to minimize the error of registration results.
作者 赵利荣 周一兵 ZHAO Lirong;ZHOU Yibing(Department of Radiotherapy,the Second Affiliated Hospital of Army Medical University,Chongqing 400037,China)
出处 《中国医疗设备》 2022年第2期47-50,共4页 China Medical Devices
基金 重庆市技术创新与应用发展专项项目(cstc2019jscxmsxmX0265)。
关键词 电子射野影像系统 感兴趣体积 盆腔仿真模体 位置验证 image verification system volume of interest pelvic simulation phantom location verification
  • 相关文献

参考文献3

二级参考文献19

  • 1罗述谦,周果宏.医学图像处理与分析[M].北京:科学出版社,2010:164-165.
  • 2Bertelsen A, Hansen CR, Johansen J, et al. Single arc volumetric mod- ulated arc therapy of head and neck cancer [J]. Radiother Oncol, 2010, 95(2): 142-148.
  • 3Clemente S, Wu B, Sanguineti G, et al. SmartArc-based volumetric modulated arc therapy for oropharyngeal cancer: A dosimetric com- parison with both intensity-modulated radiation therapy and helical tomotherapy[J]. Int J Radiat Oncol Biol Phys, 2011, 80(4) :1248-1255.
  • 4Doomaert P, Verbakel WF, Bieker M, et al. RapidArc planning and delivery in patients with locally advanced head-and-neck cancer undergoing chemoradiotherapy [J]. Int J Radiat Oncol Biol Phys, 2011, 79(2): 429-435.
  • 5Verbakel WF, Cuijpers JP, Hoffmans D, et al. Volumetric intensity mo- dulated arc therapy vs. conventional IMRT in head-and-neck cancer: a comparative planning and dosimetric study[J]. Int J Radiat Oncol Biol Phys, 2009, 74(1): 252-259.
  • 6Pluim J, Maintz J, Viergever MA. Mutual information-based registra- tion of medical images: A survey[J]. IEEE Transaction on Medical Imaging, 2003, 22(8): 986-1004.
  • 7Jeffrey Tsao. Interpolation artifacts in multimodality image registration based on maximization of mutual information [J]. IEEE Transactions on Medical Imaging, 2003, 22(7): 235-260.
  • 8Mallat S, Zhong S. Characterization of signals f:om multiscale edges [J]. IEEE Transactions on Pattern Analysis and Machine Intelligence, 1992, 14(7) : 710-732.
  • 9徐寿平,王连元,戴相昆,黄昊,解传滨.螺旋断层放疗系统原理及其应用[J].医疗卫生装备,2008,29(12):100-102. 被引量:31
  • 10尚进,李东.TOMOTHERAPY螺旋断层放射治疗的现状与应用前景[J].中国医疗设备,2012,27(2):48-50. 被引量:14

共引文献24

同被引文献19

引证文献1

二级引证文献4

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部