摘要 目的为测量光学相干断层成像(OCT)设备的检测性能提供参考。方法结合OCT设备国际标准设计模拟眼,并通过临床OCT设备验证该模拟眼用于规范测量的可行性。结果成功设计具有与正常人类视网膜类似的体积形态、散射特性和吸收特性的简洁模拟眼,并验证其用于测量OCT设备的横向分辨率、轴向分辨率、视场角、信噪比、成像深度及眼底图像和OCT扫描图像匹配等6种关键指标。结论目前市场上尚无模拟眼球的仿体可用于检测OCT设备的关键指标。制造商可结合产品特点,选择该模拟眼作为产品研发和日常质量控制。但要用作第三方检测,尚需提供更多验证数据支持。 Objective To provide a reference for measuring the performance detection of optical coherence tomography(OCT)equipment.Methods The simulated eye was designed according to the international standard of OCT equipment,and the feasibility of the simulated eye for standardized measurement was verified by clinical OCT equipment.Results A simple simulated eye with similar volume morphology,scattering characteristics and absorption characteristics as normal human retina was successfully designed,and it was verified that it could be used to measure six key indexes of OCT equipment,such as transverse resolution,axial resolution,field angle,signal-to-noise ratio,imaging depth and matching between fundus image and OCT scanning image.Conclusion At present,there is no imitation of the eyeball in the market,which can be used to measure the key indexes of OCT equipment.The manufacturer can select the analog eye for product research and development and daily quality control in combination with product characteristics.However,more verification data need to be provided before being used as a third-party test.
作者 赵荻 郭艳婧 胡毅成 侯立新 ZHAO Di;GUO Yanjing;HU Yicheng;HOU Lixin(Chongqing Institute for Food and Drug Control,Chongqing,China 401120;Chongqing Center for Medical Equipment Evaluation and Inspection,Chongqing,China 401120;Department of Biomedical Engineering,School of Future Technology,Peking University,Beijing,China 100091)
出处 《中国药业》 CAS 2022年第4期53-58,共6页 China Pharmaceuticals
关键词 光学相干断层成像 模拟眼 设备关键指标 性能检测 optical coherence tomography simulated eye key indexes of the equipment performance detection
  • 相关文献

参考文献3

二级参考文献25

  • 1Wong TY, Knudtson MD, Klein R, et al. A prospective cohort study of retinal arteriolar narrowing and mortality. Am J Epidemiol, 2004, 159: 819-825.
  • 2Hubbard LD, Brothers RJ, King WN, et al. Methods for evaluation of retinal microvascular abnormalities associated with hypertension/sclerosis in the atherosclerosis risk in communities study. Ophthalmology, 1999, 106: 2269-2280.
  • 3Wong TY, Mitchell P. Hypertensive retinopathy. N Engl J Med,2004, 351: 2310-2317.
  • 4Couper DJ, Klein R, Hubbard LD, et al. Reliability of retinal photography in the assessment of retinal microvascular characteristics:The Atherosclerosis Risk in Communities Study. Am J Ophthalmol,2002, 133:78-88.
  • 5Sherry LM, Wang JJ, Rochtchina E, et al. Reliability of computerassisted retinal vessel measurement in a population. Clin Exp Ophthalmol, 2002, 30: 179-182.
  • 6Wong TY, Klein R, Sharrett AR, et al. The prevalence and risk factors of retinal microvascular abnormalities in older people: the Cardiovascular Health Study. Ophthalmology, 2003, 110: 658-666.
  • 7Sharrett AR, Hubbard LD, Cooper LS, et al. Retinal arteriolar diameters and elevated blood pressure: The Atherosclerosis Risk in Communities study. Am J Epidemiol, 1999, 150: 263-270.
  • 8Wong TY, Klein R, Klein BEK, et al. Retinal vessel diameters and their association with age and blood pressure. Invest Ophthalmol Vis Sci, 2003, 44: 4644-4650.
  • 9Leung H, Wang J J, Rochtchina E, et al. Relationships between age, blood pressure, and retinal vessel diameters in an older population. Invest Ophthalmol Vis Sci, 2003,44: 2900-2904.
  • 10Wong TY, Hubbard LD, Klein R, et al. Retinal microvascular abnormalities and blood pressure in older people: the Cardiovascular Health Study. BrJ Ophthalmol, 2002, 86: 1007-1013.

共引文献22

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部