能谱CT虚拟平扫联合多模型迭代重建算法在甲状腺图像质量及辐射剂量中的应用 被引量:13

Adaptive Statistical Iterative Reconstruction-V Algorithm Combined with Virtual Non-Contrast Scans in Thyroid Image Quality and Radiation Dose
下载PDF
导出
摘要 目的探讨能谱CT虚拟平扫(VNC)联合多模型迭代重建算法(ASiR-V)在甲状腺图像质量及辐射剂量应用中的可行性。资料与方法收集需进行甲状腺增强检查的25例患者,采取非能谱平扫及能谱增强扫描,检查结束后在静脉期重建不同权重ASiR-V(0、20%、40%、60%、80%、100%),并重建不同权重ASiR-V的VNC,分别记作A1~A6。平扫按照滤波反投影重建,记作B1;记录各期检查辐射剂量,测量并计算6组图像感兴趣区的CT值、图像噪声(SD)及信噪比(SNR),并对图像主观和客观指标进行分析。结果A1、B1两组图像中背景噪声SD值(SD脂肪)、SNR比较,差异均无统计学意义(t=−0.196、−0.778,P均>0.05),B1图像中甲状腺CT值、SD值均高于A1图像(t=−9.206、−6.970,P均<0.05)。A1~A6图像中甲状腺SD值、SNR比较,差异均有统计学意义(F=63.258、3.987,P<0.05),随着迭代强度增大,图像SD值逐渐减小(R2=0.598,P<0.05);A1~A6图像中甲状腺CT值差异无统计学意义(F=0.130,P>0.05);A1~A6图像的0~60%主观评分呈上升趋势,40%~60%主观评分最佳(Kappa=0.83,P<0.05),80%~100%图像评分呈下降趋势,与60%图像比较差异有统计学意义(Z=−3.60、−5.47,P<0.05),权重60%的VNC主观评分高于真实平扫(Z=−4.171,P<0.05)。应用能谱虚拟平扫代替常规平扫后25例检查者的有效剂量可降低47.9%,其中静脉期能谱扫描与非能谱的有效剂量分别为(1.66±0.23)mSv、(1.75±0.24)mSv,差异有统计学意义(t=5.261,P<0.05)。结论甲状腺静脉期VNC可应用于临床诊断,降低辐射剂量47.9%;且联合后置60%权重能明显降低图像噪声,改善图像质量。 Purpose To explore the feasibility of adaptive statistical iterative reconstruction-V(ASiR-V)combined with virtual non-contrast scan(VNC)in thyroid image quality and radiation dose in the clinical application.Materials and Methods Twenty-five patients prepared for CT thyroid enhanced scan were prospectively collected.All patients received conventional true non-contrast(TNC)scanning and spectral enhanced scanning.The venous phase used spectral mode scans was used to reconstruct VNC images with different ASiR-V levels(0,20%,40%,60%,80%,100%),which were recorded as A1-A6 group,respectively.Plain scan was reconstructed by filtered back projection and recorded as B1.Radiation dosage of every phase was recorded,and the CT value,image noise(SD)and signal noise ratio(SNR)of region of interest of six groups were measured and compared,respectively.Results There was no significant difference in image background noise(SDfatty)and SNR between A1 group and B1 group(t=−0.196 and−0.778,all P>0.05).CT thyroid value and SD thyroid of B1 group image were significant higher than that of A1 group(t=−9.206,−6.970,all P<0.05).There were all statistically significant differences in image noise and SNR among A1-A6 groups(F=63.258,3.987,all P<0.05).With the increase of ASiR-V level,the image noise gradually decreased(R2=0.598,P<0.05).There was no significant difference in CT thyroid value among A1-A6 groups.With the increase of ASiR-V level,the subjective scores of images showed gradually increases.Among A1-A6 groups,the subjective score of 40%-60%ASiR-V image was the best(Kappa=0.83,P<0.05).The subjective score of 80%-100%ASiR-V image showed a downward trend,and the difference was statistically significant compared with that of 60%ASiR-V image(Z=−3.60,−5.47,P<0.05).The subjective score of VNC image with 60%ASiR-V image was significantly higher than that of TNC image.The effective dose(ED)of 25 patients was decreased by 47.9%after using energy spectrum virtual scanning replacing conventional scanning.The ED of venous energy spectrum scanning and non-energy spectrum scanning were(1.66±0.23)mSv and(1.75±0.24)mSv,respectively,and there was significant difference in the ED among A1-A6 groups(t=5.261,P<0.05).Conclusion VNC scan of thyroid venous phase can meet the requirements of imaging diagnosis,reducing radiation dose by 47.9%.VNC scan combined with 60%ASiR-V in the vein phase of thyroid energy spectrum scan can significantly reduce radiation dose and improve image quality in clinical application.
作者 田兆荣 孙杰 杨建平 倪亚博 石晓萌 沈云 陈泳 王志军 TIAN Zhaorong;SUN Jie;YANG Jianping;NI Yabo;SHI Xiaomeng;SHEN Yun;CHEN Yong;WANG Zhijun(Department of Radiology,General Hospital of Ningxia Medical University,Yinchuan 750001,China)
出处 《中国医学影像学杂志》 CSCD 北大核心 2022年第2期129-133,138,共6页 Chinese Journal of Medical Imaging
基金 宁夏医科大学校级课题(XM2019100) 2021宁夏回族自治区重点研发项目(2021BEG03033)。
关键词 甲状腺 体层摄影术 X线计算机 迭代重建技术 算法 图像处理 计算机辅助 辐射剂量 质量控制 Thyroid gland Tomography,X-ray computed Iterative reconstruction Algorithms Image processing,computer-assisted Radiation dosage Quality control
  • 相关文献

参考文献4

二级参考文献33

  • 1吴晓华,马大庆,张忠嘉,冀景玲,张岩松.多层螺旋CT胸部低剂量扫描发现肺结节的临床研究[J].中华放射学杂志,2004,38(7):767-770. 被引量:86
  • 2王晓峰,彭明辰.关于多排螺旋CT剂量的探讨[J].医疗设备信息,2004,19(12):23-25. 被引量:20
  • 3殷瑞根,吴树春,王冬青,彭卫斌,孙维斌,范志刚,廖谦,赵亮.低剂量多层螺旋CT在急腹症中的初步应用[J].临床放射学杂志,2005,24(7):609-612. 被引量:29
  • 4头颈部CT、MR扫描规范指南(修改稿)[J].中华放射学杂志,2007,41(9):996-999. 被引量:41
  • 5Barrett J F, Keat N. Artifacts in CT : recognition and avoidance. Ra- dioGraphics,2004,24 : 1679.
  • 6Brink M, de Lange F, Oostveen LJ, et al. Arm raising at exposure controlled multi-detector trauma Ctof thoracoabdominal region : higher imagequality, lower radiation dose. Radiology, 2008,249:661.
  • 7Mayo JR, Hertman TE, lee KS, et al. CT of the chest : minimaI tube current required for good image quaIity with the least radiation dose. A JR, 1995,164:603.
  • 8Mullins ME, Lev MH, Bove P, at al. Comparison of image quality be- tween conventional and lowdose nonenhanced head CT. AJNR, 2004,25:533.
  • 9Sohaib SA,Peppercorn PD,Horrocks JA, et al. The effect of decrea- sing mAs on image quality and patient dose in sinus CT. B JR Br J Radiol,2001,74 : 157.
  • 10Brenner D J, Elliston CD, Hall EJ, et aI. Estimated risks of radiation- induced fatal cancer from pediatric CT. AJR ,2001,176:289.

共引文献49

同被引文献146

引证文献13

二级引证文献10

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部