A scintillator detector for beam tuning of low-energy single electron accelerator

原文传递
导出
摘要 Purpose Nine scintillator detectors were developed for beam tuning of low-energy single electron accelerator instead of the picoammeter and Faraday cup,both of which will be insensitive to the weak-current beam lower than picoampere.Methods The detectors are based on plastic scintillator and photomultiplier tube(PMT)and the beam is 100 Hz pulsed electrons in the energy range of 0.1 to 50MeV.The beam flux was tuned from about 107 to quasi-single electron for each bunch according to the measurement results of these detectors installed along the beam pipe and after slits.In order to determine the number of electrons in a bunch,the gains of each PMT were finely calibrated,and the energy response of each detector was calibrated with 207Bi radioactive source.Moreover,the deposited energy in the plastic scintillator,the electron detection efficiency and light collection efficiency,as a function of electron incident energy,were also simulated in detail by Geant4.Results Nine detectors worked well at a gain of 0.6×10^(6) after installation in the beam pipe,and the number of electrons in a bunch could be measured at each detector position.The single electron peak was seen when the beam reached the final detector.Conclusion The commissioning result shows that the beam was tuned to quasi-single electron level successfully and the scintillator detector is an effective method to measure the weak-current beam lower than picoampere.
出处 《Radiation Detection Technology and Methods》 2017年第1期29-39,共11页 辐射探测技术与方法(英文)
基金 This work is also supported byNationalNatural Science Foundation of China(10565004) We acknowledge the support and cooperation of Hainan Zhanchuang Photonics Technology Co.,Ltd and Anderson Superconducting RF Accelerator Technology INC.We thank the members of the Beijing Test Beam Facility(BTBF)for their help.
  • 相关文献

参考文献3

二级参考文献37

  • 1BESⅢ Preliminary Design Report.http://epc.ihep.ac.cn/epc/ design/ design 1.htm
  • 2Vasil'chenko V G,Lapshin V G,Peresypkin A I et al.Nucl.Instrum.Methods,1996,A369:55
  • 3Hamada M M,Rela P R,Costa F E et al.Nucl.Instrum.Methods,1999,A422:148
  • 4Majewski S,Bowen M,Johnson C Z et al.Nucl.Instrum.Methods,1989,A281:500
  • 5Bodmann B,Goeb S,Holm U.Nucl.Instrum.Methods,2003,B208:495
  • 6Aschenauer E C,Baehr J,Nahnhauer R et al.arXiv:physics/9907018vl
  • 7Zorn C,Bowen M,Majewski S et al.Nucl.Instrum.Methods,1989,A276:58
  • 8Marini C,Donatelli I,Nigro A et al.Radiation Damage to Organic Scintillation Materials:CERN 85-08.June 1985
  • 9Britvich G I,Peresypkin A I,Rykalin V I et al.Nucl.Instrum.Methods,1993,A326:483
  • 10http://www.detectors.saint-gobain.com/Media/Documents/S30000000000000000003/organics-brochure-2001.pdf

共引文献4

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部