[目的]比较单髁关节置换运动学对线定位技术与传统手术技术,在术中垫片运动轨迹、临床效果和影像学评价上的差异.[方法]回顾性分析2019年1-6月收治牛津单髁手术患者.其中,23例(30膝)术中采用运动学对线技术(运动学组),另外25例(30膝)术中采用常规对线技术(常规组),两组患者一般资料差异无统计学意义(P>0.05).比较两组术中垫片运动方式和运动距离、临床效果及影像学资料.[结果]术中垫片运动轨迹存在差异,运动组可达到更为理想的垫片运动轨迹(P<0.05),而术中垫片移动距离差异无统计学意义(P>0.05).临床结果显示,运动组手术时间和出血量均显著少于常规组(P<0.05).相应时间点,两组间KSS评分差异均无统计学意义(P>0.05).影像评估显示,两组患者术后的髋膝踝角(hip-knee-ankle angle,HKAA)均较术前显著增加(P<0.05),相应时间点两组间HKAA的差异均无统计学意义(P>0.05).术后两组间股骨及胫骨假体内外翻、股骨假体屈伸、胫骨假体后倾、假体相对倾斜角比较差异均无统计学意义(P>0.05).但是,运动组的假本邻近度显著小于常规组(P<0.05).[结论]本研究中介绍的膝关节单髁置换术(unicompartmental knee arthroplasty,UKA)中运动学对线定位技术与传统手术相比,临床效果相似,假体安装准确性相似,但手术时间更短、术中出血量更小、假体间的邻近度更小,且垫片运动轨迹更为理想.
[Objective]To clarify whether the kinematic alignment technique improve the trajectory of the bearing movement,as well as the clinical outcomes and radiological presentations in Oxford mobile bearing unicompartmental knee arthroplasty(Oxford MB UKA).[Methods]We retrospectively analyzed the patients who underwent UKA from January to June,2019.Of them,23 patients(30 knees)had UKA performed by the kinematic alignment technique(the kinematic group),while the other 25 patients(30 knees)received UKA with the conventional alignment technique(the conventional group).There were no significant differences in term of baseline data preoperatively between the two groups.The intraoperative bearing movement mode,motion distance,clinical outcome and radiographic data were compared between the two groups.[Results]The kinematic group proved significantly superior to the conventional group in term of bearing movement trajectory(P<0.05),despite of the fact that no a difference was noticed in the motion distance of the liner between them(P>0.05).In terms of clinical outcomes,the kinematic group was significantly superior to the conventional group regarding to intraoperative blood loss and operation time(P<0.05),whereas there were no statistically significant differences in KSS clinical and functional scores between the two groups at any matching time point(P>0.05).Radiographically,the hip-knee-ankle angle(HKAA)significantly decreased postoperatively compared with that preoperatively in both groups(P<0.05),which was not statistically different between the two groups at any corresponding time point(P>0.05).There were no significant differences in terms of varus-valgus angles of the femoral and tibial components,flexion-extension angle of the femoral component,posterior slope of the tibial tray and relative oblique of the femoral and tibial components between the two groups(P>0.05).However,the kinematic group got significantly closer contiguity of the components than the conventional group(P<0.05).[Conclusion]Although the kinematic technique does achieve similar prosthetic installation accuracy and clinical outcome with the conventional technique for Oxford MB UKA,the former has advantages of less blood loss,shorter operation time,closer contiguity of the femoral and tibial components,and more optimal bearing movement trajectory over the latter.
作者
孙晓威
张启栋
王卫国
程立明
郭万首
SUN Xiao-wei;ZHANG Qi-dong;WNAG Wei-guo;CHENG Li-ming;GUO Wanshou(Department of Orthopaedics,China-Japart Friendship Hospital,Beijing 100029,China;Postgraduate School,Peking Union Medical College,Beijing 100029,China)
出处
《中国矫形外科杂志》
CAS
CSCD
北大核心
2022年第3期198-202,共5页
Orthopedic Journal of China
关键词
单髁置换术
运动学对线技术
垫片运动轨迹
手术技术
unicompartmental knee arthroplasty
kinematic alignment technique
bearing movement trajectory
surgical technique