基于掺铥光纤激光器的体外碎石实验探究 被引量:1

Exploration of Thulium-Doped Fiber Lasers in Lithotripsy in vitro
原文传递
导出
摘要 采用线型腔结构,实现了中心波长为1939.31 nm、脉宽在0~2000μs可调、重复频率在0~2 kHz可调、最大平均输出功率为34.2 W的准连续掺铥光纤激光器,并利用此激光器在体外环境下开展结石消融量随激光脉宽与重复频率的变化规律的研究。结果显示:在相同时间内,当单脉冲能量相近时,增大脉冲重复频率(平均输出功率)有利于提高碎石速率;当平均输出功率接近时,单脉冲能量越大,石块消融量越大。在90 s的碎石时间内,石块经过31.8 W/0.053 J(250μs)、33.1 W/0.11 J(500μs)、33.5 W/0.22 J(1000μs)、34.2 W/0.45 J(2000μs)4组参数激光照射后的消融量分别为0.333,0.480,0.697,0.723 g,结石表面的最高水温分别为30.8,35.5,38.9,41.2℃。 Objective Water has a strong absorption coefficient in the mid-infrared band,which explains why mid-infrared lasers are widely used in medicine.The thulium-doped fiber laser(TFL)emission wavelength range includes the water absorption peak at 1940 nm,which has been used in prostatic hyperplasia tissue resection surgery.There have been numerous reports on TFL being used in lithotripsy,including the mechanism of action and clinical tests.There was,however,no mention of a long pulse mode or a high pulse repetition rate mode.We developed a quasicontinuous wave(QCW)TFL for laser lithotripsy to address these issues.In conjunction with previous reports,this study focuses on the use of various laser parameters to ablate calculus in vitroin order to verify the feasibility of TFL laser lithotripsy and evaluate calculus ablation efficiency.Additionally,a thermometer is used to monitor the peak temperature of the water to ensure that the system is operating safely.Methods Six 140 W,793 nm fiber-coupled laser diodes(LDs)pumped the QCW TFL,with the six LDs modulated synchronously by the drive current signal.The pulse width and repetition rate(PRR)of the LD drive current can be continuously tuned in the range of 0-2 ms and 0-2 kHz.Two fiber Bragg gratings(FBG)were used to construct the reflectance of laser oscillator.The reflectance of high-reflective FBG is 99.8%,while the reflectance of low-reflective FBG is 9.38%.The gain fiber was a 4-m-long thulium-doped fiber with a core diameter of 25μm and an inner cladding diameter of 400μm.At 793 nm,the cladding absorption is 4.2 dB/m.The laser output port was a SMA905 connector that could be butt joined to a medical fiber via a converter.To simulate the clinical environment for laser lithotripsy,the calculus was placed in water,as shown in Fig.4.Calcium sulfate was used as calculus in the experiment to investigate the effect of TFL parameters on calculus ablation efficiency and ambient temperatures.Three times before and after laser exposure,the mass loss of the calculus was determined.Finally,a thermometer was used to monitor the temperature of the water at a depth of 3 mm below the surface of the calculus.Results and Discussions The TFL laser had a wavelength of 1939.31 nm and a maximum average output power of34.2 W.After modulation,the single pulse profiles and pulse trains were stable(Fig.3).The experimental phenomena of laser lithotripsy revealed that when the medical fiber tip struck the surface of a calculus,it immediately produced a large amount of powder that diffused into the water,and the calculus’ s surface had some burnt marks.Consequently,we conclude that there is an obvious photo thermal ablation mechanism in the process of TFL laser lithotripsy in a water environment.Table 2 displayed the measured mass loss of the calculus with various laser parameters.Mass loss increased by 0.09 g(200 Hz/9.53 W/0.047 J)up to 0.29 g(400 Hz/20.29 W/0.05 J)and 0.333 g(600 Hz/31.8 W/0.053 J)at a pulse width of 250μs.At first,the mass loss increased with increasing output power at pulse widths of 500,1000,and 2000μs.Consequently,whether the laser pulse is long or short,the ablating rate increases with increasing laser power.Furthermore,the mass loss of ablation increased with larger single pulse energy while output power remained approximate,such as 0.333 g at 31.8 W/0.053 J(250μs),0.480 g at 33.1 W/0.11 J(500μs),0.697 g caused by 33.5 W/0.22 J(1000μs),and 0.723 g at 34.2 W/0.45 J(2000μs).In this case,the single pulse energy was the primary influencing factor in calculus ablation.Moreover,the single pulse energy had an effect on not only ablation efficiency but also water temperature.When single pulse energy reached 0.45 J(2000μs/34.2 W)during laser irradiation,the peak temperature of water was 41.2℃.As a result,for the same output power,the single pulse energy produced more heat in the same amount of time.Conclusions QCW TFL was constructed with a center wavelength of PRR of 1939.31 nm,apulse width of 0-2000μs,and a PRR of 0-2 kHz,respectively.34.2 W was the maximum average output power.Following that,in vitro lithotripsy experiments were performed using the laser.The results indicate that when the single pulse energy is constant,the average output power of the laser is the primary factor affecting the ablation efficiency of lithotripsy.And when the average output power is comparable,the energy of the single pulse becomes the determining factor.Additionally,the TFL can generate more heat when operating in the long pulse width mode,resulting in a greater rise in water temperature.In conclusion,TFL with a wavelength of 1940 nm has a significant lithotripsy effect.When combined with fiber lasers’ superior characteristics,it is very likely to become the laser source for the next generation of laser lithotripsy.
作者 林宇 刘敏秋 欧阳德钦 肖克峰 陈业旺 吕启涛 阮双琛 Lin Yu;Liu Minqiu;Ouyang Deqin;Xiao Kefeng;Chen Yewang;LüQitao;Ruan Shuangchen(Key Laboratory of Advanced Optical Precision Manufacturing Technology of Guangdong Higher Education Institutes,Shenzhen Technology University,Shenzhen,Guangdong 518118,China;Sino-German College of Intelligent Manufacturing,Shenzhen Technology University,Shenzhen,Guangdong 518118,China;Shenzhen People's Hospital,Shenzhen,Guangdong 518020,China)
出处 《中国激光》 EI CAS CSCD 北大核心 2022年第1期171-179,共9页 Chinese Journal of Lasers
基金 国家自然科学基金(61905146) 深圳市科技计划基础研究(自由探索)项目(JCYJ20180301171044707)。
关键词 激光光学 激光碎石 掺铥光纤激光器 准连续激光器 结石消融 laser optics laser lithotripsy thulium-doped fiber lasers QCW lasers calculus ablation
  • 相关文献

参考文献3

二级参考文献35

  • 1Teng P,Nishioka NS,Anderson RR,et al.Optical studies of pulsed laser fragmentation of biliary calculi[J].Appl Phys B,1987,42:73-78.
  • 2Teng P,Nishioka NS,Anderson RR,et al.Acoustic studies of the role of immersion in plasma-mediated laser ablation[J].IEEE J Quantum Electron,1987,23:1845-1852.
  • 3Lo EY,Petschek H,Rosen DI.A hydrodynamic model for the laser induced fragmentation of calculi[J].Laser Life Sci,1990,3:233-244.
  • 4Steven AS,Frederic MD,Babaeddin J,et al.Mechanisms of biliary stone fragmentation using the Ho:YAG laser[J].IEEE Trans Biomed Eng,1994,41:276-283.
  • 5Malcom S,Keith R.Determination of the shock wave intensity from a laser lithotriptor using a bi-laminar hydrophone[C].Innovative Pressure,Force and Flow Measurements,IEE Colloquium on,1999.10:1-5.
  • 6Rink K,Delacrétaz G,Salathé RP,et al.Fragmentation process induced by microsecond laser pulses during lithotripsy[J].Appl Phys Lett,1992,61:258-260.
  • 7Olmes A,Franke HG,Bansch E,et al.Modeling of infrared soft-tissue photoablation process[J].Appl Phys B,1997,65:659-666.
  • 8Hibst R,Keller U.Experimental studies of the application of the Er:YAG laser on dental hard substances.Ⅰ.Measurement of the ablation rate[J].Lasers Surg Med,1989,9:338-344.
  • 9Ostertag M,McKinley JT,Reinisch L,et al.Laser ablation as a function of the primary absorber in dentin[J].Lasers Surg Med,1997,21:384-394.
  • 10Chan KF,Choi B,Vargas G,et al.Free electron laser ablation of urinary calculi:an experimental study[J].IEEE J Select Topics Quantum Electron,2001,7:1022-1032.

共引文献12

同被引文献14

引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部