血细胞分离机单采过程抗凝剂和生理盐水连接错误原因分析及预防

Cause analysis and prevention of wrong connection between anticoagulant and normal saline solution during apheresis platelet donation
下载PDF
导出
摘要 目的探讨血细胞分离机单采过程抗凝剂和生理盐水错误连接的原因及有效预防措施,保障单采献血者的安全。方法第1阶段2008年12月至2016年9月采用人工校核方法;第2阶段2016年10月至2020年10月采用人工和信息系统辅助双校核方法。行Fisher’s Exact Test双侧检验,对比校核方法改进前后溶液连接错误发生率。结果第1阶段单采过程溶液连接错误发生率为1.02/万;Amicus与Trima+Mcs;+血细胞分离机溶液连接差错率之间,P<0.05;第1、2阶段总差错率比较,P>0.05;第2阶段推行人工和信息系统双校核方法后,未再发生抗凝剂和生理盐水连接错误的情况。结论采用人工和信息系统辅助双校核方法,可有效预防抗凝剂和生理盐水连接错误的发生。 Objective To explore the reasons for wrong connection between anticoagulant and normal saline solution during apheresis platelet donation, as well as the preventive measures, so as to ensure the safety of apheresis platelet donors. Methods Manual checking in the first phase(December 2008 to September 2016) was compared with double checking(manual checking plus information system) in the second phase(October 2016 to October 2020) via bilateral testing using Fisher’s Exact Test to study pre-post-improvement differences. Results The incidence of solution connection errors during apheresis platelet donation in the first phase was 1.02/10 000, and the error incidence between Amicus and Trima + Mcs;blood cell separator was statistically significant(P<0.05). The total incidence of errors between the first and second phases was not statistically significant(P>0.05). After the performance of double checking in the second phase, no wrong connection of anticoagulant and saline solution occurred. Conclusion The double checking method assisted by manual and information system can effectively prevent the wrong connection between anticoagulant and normal saline solution.
作者 纪淑华 黄淑铭 汤辉伟 陈岑 褚晓凌 JI Shuhua;HUANG Shuming;TANG Huiwei;CHEN Cen;CHU Xiaoling(Fujian Blood Center,Fuzhou 350004,China.)
机构地区 福建省血液中心
出处 《中国输血杂志》 CAS 2022年第2期219-221,共3页 Chinese Journal of Blood Transfusion
关键词 单采血小板 溶液连接 信息系统 辅助校核 apheresis platelets solution connect information system auxiliary calibration
  • 相关文献

参考文献7

二级参考文献56

  • 1ISO 80369-1:2011 (E), Small-bore connectors for liquids and gases in healthcare applications - Part 1: General requirements.
  • 2D. Simmons & K. Graves, Luer Tip (Small Bore Medical Connector) Misconnection Reference List, UT M. D. Anderson Cancer Center. 4/25/2008.
  • 3U.S. Inaction Lets Look-Alike Tubes Kill Patients, available at: http://www.nytimes.com/2010/08/21/health/ policy/21 tubes.html?_r=2.
  • 4SUZANNE C. BEYEA, Caution: Tubing misconnections can be deadly. AORN JOURNAL MARCH,2007, 85(3):633-635.
  • 5Tubing misconnections - a persistent and potentially deadly occurrence, Sentinel Event Alert, The Joint Commission, 3 April 2006 http://www.jointcommission. org/SentinelEvents/Sentinel EventAlert/sea_36.htm.
  • 6Groups Work to Eliminate Medical Errors From Tube Misconnections, Medical News Today, 29 June 2007 http://www.medicalnewstoday.com/articles/75466.php.
  • 7Problems Persist with Life-Threatening Tubing Misconnections, ISMP Medication Safety Alert, 17 June 2004 http://www.ismp.org/newsletters/acutecare/ articles/20040617.asp.
  • 8GB18671-2002,一次性使用静脉输液针.
  • 9GB18671-2009,一次性使用静脉输液针.
  • 10GB/T1962.1-2001,注射器、注射针及其他医疗器械6%(鲁尔)圆锥接头第1部分:通用要求.

共引文献40

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部