心磁图仪系统性能测试方法研究

Research on the System Performance Test Method for MCG
下载PDF
导出
摘要 心磁图仪作为一种新型医疗技术,其临床应用效果显著。为了可靠地检测心磁图仪产品,本文提出了受测者定位精度、磁场-电压转换系数、通道差异率、磁灵敏度、动态范围和系统采样率5项性能指标及其检测方法。这些指标有助于规范心磁图仪产品标准,保证其临床应用安全性和有效性。 Magnetocardiography(MCG) is a new medical technology with remarkable effect of clinical application. In order to reliably detect the magnetocardiography, 5 performance indexes and their detection methods are proposed in this paper such as positioning accuracy of the subject, the magnetic field voltage conversion coefficient, the channel difference rate, the magnetic sensitivity, the dynamic range and the system sampling rate. These indexes are helpful to specialize the product standard of MCG and ensure its safety and effectiveness in clinical application.
作者 艾海明 郑晓雷 吕卓然 李勇 AI Hai-ming;ZHENG Xiao-lei;LYU Zhuo-ran;LI Yong(Faculty of Science and Technology,Beijing Open University;University of California,Davis;Beijing Milestone Science&Technology Development Co.,Ltd.)
出处 《标准科学》 2022年第3期109-112,共4页 Standard Science
基金 北京市教委科技项目(项目编号:KM201951160001) 2019年海淀区支持城市可持续发展和社会事业研发专项“心磁图仪系统开发与临床应用研究”资助。
关键词 心磁图仪 心磁信号 临床应用 系统性能 测试方法 magnetocardiography(MCG) magnetocardiosignal clinical application system performance test method