高剂量率铱源剂量学特征的蒙特卡罗模拟研究 被引量:2

Dosimetric Characteristics Study on the Ir-192 High Dose Rate by Monte Carlo Simulation
下载PDF
导出
摘要 目的采用蒙特卡罗模拟技术,借助程序MCNP5对国产高剂量率铱源剂量率的分布进行模拟研究。方法基于TG43号报告中的剂量率计算公式,借助MCNP5建立铱源模型,模拟计算此模型的空间剂量分布,分析研究铱源的径向剂量函数及各向异性函数。结果模拟的各向异性函数与发表结果在误差范围内一致。当径向距离较小时,模拟的径向剂量函数与发表结果符合较好;当径向距离较大时,模拟的径向剂量函数与部分发表结果符合较好,但也与部分发表结果相差较大,在r=15 cm处,最大相对误差为24%。结论基于MCNP5的蒙特卡罗模拟技术可较好地模拟研究国产高剂量率铱源的剂量学特征,这为进一步探索影响后装放疗剂量分布的因素奠定了理论基础。 Objective To simulate the dose rate distribution of domestic high dose rate Ir-192 with the help of the program MCNP5 using Monte Carlo simulation technology.Methods Based on TG43 protocol,the Ir-192 modelling was established through MCNP5 code.At then,the distribution of dose rate in space was simulated with this modeling.From the simulation,the radial dose function and anisotropy function were obtained.Results The simulated anisotropy function was in agreement with the published results within the error range.When the radial distance was small,the radial dose function agreed with each other in the small radial distance;for the big radial distance,the radial dose function was only agree with some published results.The largest relative difference was 24%at radial distance 15 cm.Conclusion The Monte Carlo technique based on MCNP5 can better simulate the dosimetry characteristics of domestic high-dose rate iridium source,which lays a theoretical foundation for further exploring the factors affecting the dose distribution of radiotherapy after loading.
作者 吴晓维 崔相利 刘子成 刘苓苓 WU Xiaowei;CUI Xiangli;LIU Zicheng;LIU Lingling(Radiotherapy Center,Hefei Cancer Hospital,Chinese Academy of Sciences,Hefei Anhui 230031,China;School of Biomedical Engineering,Anhui Medical University,Hefei Anhui 230032,China)
出处 《中国医疗设备》 2022年第4期6-9,共4页 China Medical Devices
基金 国家重点研发计划(2018YFE0114100) 安徽省重点研究与开发计划(1804h08020279,202004j07020052)。
关键词 铱源剂量学特征 蒙特卡罗模拟 高剂量率 铱-192 dosimetric characteristics of iridium source monte carlo simulation high dose rate Ir-192
  • 相关文献

参考文献12

二级参考文献90

  • 1郁志龙,索志敏,阎瑞敏,张哲林,张保祯,牛广明.240例放射治疗宫颈癌回顾性分析[J].内蒙古医学杂志,2004,36(10):786-788. 被引量:11
  • 2鲁玮瑗,李智华,黄菊英,李延玲,吴瑞.近距离后装治疗机及其在鼻咽癌治疗中的应用[J].医疗设备信息,2006,21(11):50-51. 被引量:8
  • 3Nag S,Orton C,Young D,et al.The American Brachytherapy Society Survey of brachytherapy practice for carcinoma of the cervix in the United States[J].Gyn Oncol,1999,73:111-118.
  • 4International Commission on Radiation Units and Measurements (ICRU).Dose and volmne specifications for reporting intraeavitary therapy in gynecology,ICRU Report no.38[R].Bethesda,MD:ICRU,1985.
  • 5Wang X,Liu R,Ma B,et.al.High dose rate versus low dose rate intracavity brachytherapy for locally advanced uterine cervix cancer[J].Cochrane Database Syst Rev,2010,7:CD007563 DOI:10.1002/14651858.
  • 6Narayan K,van Dyk S,Bemshaw D,et.al.Comparative study of LDR (Manchester system)and HDR image-guided conformal brachytherapy of cervical cancer:patterns of failure,late complications,and survival[J].Int J Radiat Oncol Biol Phys,2009,74(5):1529-1535.
  • 7Boss E,Barentsz J,Massugen L,et al.The role of MR imaging in invasive cervical carcinoma[J].Eur Radiol,2000,10:256-270.
  • 8Viswanathan AN,Dimopoulos J,Kirisits C,et al.CT-versus MRIbased contouring in cervical cancer brachytherapy:results of a prospective trial and preliminary guidelines for standardized contours[J].Int J Radiat Oncol Biol Phys,2007,68:491-498.
  • 9Khoo VS,Dearualey DP,Finnigan DJ,et.al.Magnetic resonance inmging (MRI):considerations and applications in radiotherapy treatment planning[J].Radiother Oncol,1997,42:1-15.
  • 10Petersch B,Bogner J,Fransson A,,et al.Effects of geometric distortion in 0.2T MRI on radiotherapy treatment planning of prostate cancer[J].Radiother Oncol,2004,71:55-64.

共引文献104

同被引文献11

引证文献2

二级引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部