摘要
以美国锐珂VX3733-SYS型和DRXR-1型数字X射线摄影系统(DR)故障为例,分析3例设备无法曝光的原因,通过逐步排查提出解决方案,为同行提供参考。
Taking failures of VX3733-SYS and DRXR-1 of Carestream Health digital X-ray photographic system(DR)as examples,the reasons for failure of exposure of three devices were described and solutions were proposed through step-by-step investigation to provide reference for the industry.
作者
车雪松
张敏
卢东生
CHE Xue-song;ZHANG Min;LU Dong-sheng(Department of Medical Engineering,Beijing Friendship Hospital,Capital Medical University,Beijing 100050,China;不详)
出处
《中国医学装备》
2022年第5期205-207,共3页
China Medical Equipment