摘要
采用海洋学中广泛使用的水中^(226)Ra浓度测定方法,进行超低浓度水样(<0.2Bq/m^(3))的测定。探讨了测定过程中的各种本底来源和降低本底的措施。采取各种措施后,残余空白本底的总和远远小于12 mBq,满足了测定浓度<0.2 Bq/m^(3)的超低浓度镭溶液的要求。提示了进一步降低本底的可能性。
The^(226) Ra concentration measurement method in water widely used in oceanography is used to measure trace concentration water samples(<0.2 Bq/m^(3)).Various background sources and measures to reduce the background in the determination process are discussed.After taking corresponding measures,the sum of the residual blank background is far less than 12 mBq,which meets the requirements for the determination of trace concentration radium solution with a concentration of<0.2 Bq/m^(3).The possibility of further reducing the background is suggested.
作者
朱少璞
丘寿康
唐泉
陈妍宇
张旭
ZHU Shao-pu;QIU Shou-kang;TANG Quan;CHEN Yan-yu;ZHANG Xu(School of Nuclear Science and Technology,Nanhua University,Hengyang Hunan 421001,China)
出处
《核电子学与探测技术》
CAS
北大核心
2021年第6期1093-1096,共4页
Nuclear Electronics & Detection Technology
基金
中国自然科学基金雅砻江联合基金(U1865208)
湖南省教育厅优秀青年基金(19B504)资助。
关键词
锰纤维
闪烁室
镭
氡
Manganese Fiber
Scintillation Cell
Radium
Radon