摘要
介绍了高性能载人离心机变频器柜ACS6000的基本结构,详细分析了变频器柜水冷单元水压低的具体原因及排除方法,指出了排除故障时的注意事项,为同行提供了参考。
The basic composition of ACS6000 frequency converter cabinet of the high performance human centrifuge were introduced. The specific cause for the low water pressure in the water cooling unit of the cabinet was analyzed in detail, and the countermeasures were put forward accordingly. Some precautions were pointed out to provide references for treating similar faults. [Chinese Medical Equipment Journal,2022,43(6):97-99,108]
作者
李毅峰
蒋科
李宝辉
王海霞
杨明浩
金朝
徐艳
王红
张立辉
杨景慧
LI Yi-feng;JIANG Ke;LI Bao-hui;WANG Hai-xia;YANG Ming-hao;JIN Zhao;XU Yan;WANG Hong;ZHANG Li-hui;YANG Jing-hui(Air Force Medical Center,PLA,Beijing 100142,China)
出处
《医疗卫生装备》
CAS
2022年第6期97-99,108,共4页
Chinese Medical Equipment Journal
关键词
载人离心机
变频器柜
水冷单元
故障维修
航空航天医学
human centrifuge
frequency converter cabinet
water cooling unit
fault maintenance
aerospace medicine