期刊文献+

载人离心机变频器柜的基本结构和故障维修

Basic composition and fault maintenance of frequency converter cabinet of human centrifuge
下载PDF
导出
摘要 介绍了高性能载人离心机变频器柜ACS6000的基本结构,详细分析了变频器柜水冷单元水压低的具体原因及排除方法,指出了排除故障时的注意事项,为同行提供了参考。 The basic composition of ACS6000 frequency converter cabinet of the high performance human centrifuge were introduced. The specific cause for the low water pressure in the water cooling unit of the cabinet was analyzed in detail, and the countermeasures were put forward accordingly. Some precautions were pointed out to provide references for treating similar faults. [Chinese Medical Equipment Journal,2022,43(6):97-99,108]
作者 李毅峰 蒋科 李宝辉 王海霞 杨明浩 金朝 徐艳 王红 张立辉 杨景慧 LI Yi-feng;JIANG Ke;LI Bao-hui;WANG Hai-xia;YANG Ming-hao;JIN Zhao;XU Yan;WANG Hong;ZHANG Li-hui;YANG Jing-hui(Air Force Medical Center,PLA,Beijing 100142,China)
出处 《医疗卫生装备》 CAS 2022年第6期97-99,108,共4页 Chinese Medical Equipment Journal
关键词 载人离心机 变频器柜 水冷单元 故障维修 航空航天医学 human centrifuge frequency converter cabinet water cooling unit fault maintenance aerospace medicine