SYSMEX XN-550全自动血液分析仪的校准及性能评价 被引量:2

Calibration and performance evaluation of the SYSMEX XN-550 fully automatic blood analyzer
下载PDF
导出
摘要 目的对SYSMEX XN-550全自动血液分析仪检测系统进行校准及性能评价.方法按照WS/T 347-2011《血细胞分析的校准指南》及WS/T 406-2012《临床血液学检验常规项目的分析质量要求》,对血细胞分析仪XN-550进行校准及性能评价,即对血细胞分析中白细胞(WBC)、红细胞(RBC)、血红蛋白(HB)、平均红细胞体积(MCV)、平均红细胞血红蛋白浓度(MCHC)、红细胞压积(HCT)、血小板(PLT)等7个主要检测项目进行本底计数、携带污染率、批内精密度、日间精密度、线性、正确度、不同吸样模式的结果可比性、实验室内结果可比性、准确度、仪器五分类比对及人员与仪器间白细胞分类计数的比较(人机五分类)等性能进行检测.结果SYSMEX XN-550全自动血液分析仪校准合格,在规定校准时效内可进行相关检测.其性能测试包括本底计数、携带污染率、批内精密度、日间精密度、线性、正确度、不同吸样模式的结果可比性、实验室内结果可比性、准确度、仪器间五分类比对、人机五分类实验等检测均通过验证.结论SYSMEX XN-550全自动血液分析仪校准合格,各分析性能良好,达到了临床血液实验室要求,且满足ISO 15189现场评审要求. Objective To calibrate and evaluate the performance of SYSMEX XN-550 automated hematology analyzer.Methods According to WS/T 347-2011 calibration guidelines for blood cell analysis and WS/T 406-2012 analysis quality requirements for routine items of clinical hematology testing,the automated hematology analyzer XN-550 was calibrated and the performance of the instrument was evaluated.The seven major evaluated items included WBC,RBC,HB,MCV,MCHC,HCT and PLT.And we evaluated these items in background count,carryover rate,between-run precision,inter-day precision,linear range,accuracy,precision,the comparability of different sample absorption mode,and the comparison of five classification between instrument and manual-counting.Results SYSMEX XN-550 automatic hematology analyzer was qualified for calibration,and relevant tests could be carried out within the prescribed calibration time.Its performance tests include background count,carryover rate,between-run precision,inter-day precision,linear range,accuracy,precision,the comparability of different sample absorption mode,and the comparison of five classification between instrument and manual-counting,and these tests all passed validation.Conclusion The SYSMEX XN-550 automatic hematology analyzer is qualified for calibration and has good analytical performance.It meets the requirements of clinical blood laboratory and meeting the requirements of IS015189 on-site review.
作者 谢楠 杨斯桀 刘丽萍 曹坤 郑大伟 王松松 高选 Xie Nan;Yang Sijie;Liu Liping;Cao Kun;Zheng Dawei;Wang Songsong;Gao Xuan(Reproductive Medicine Research Center of Shandong University,National Research Center for Assisted Reproductive and Eugenics Engineering Technology,Key Laboratory of the Ministry of Reproductive Endocrinology Department of Shandong University,Jinan 250001,China)
出处 《现代仪器与医疗》 CAS 2022年第2期5-9,共5页 Modern Instruments & Medical Treatment
关键词 全自动血液分析仪 校准 性能评价 Fully automatic blood analyzer Calibration Performance evaluation
  • 相关文献

参考文献2

二级参考文献4

共引文献13

同被引文献15

引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部