摘要 为了提供更便携和人机交互性更友好的光学相干断层成像(Optical coherence tomography,OCT)控制系统,通过分析频域OCT系统在不同扫描方式下所需的控制时序,基于ARMv7-ME架构的芯片STM32F407ZGT6和嵌入式操作系统μC/OS-Ⅲ,提出了一种用于检测人眼的嵌入式的频域OCT成像控制系统设计方案。从硬件和软件方面,详细描述了嵌入式OCT成像控制系统的构成与运行的过程。最后通过使用嵌入式OCT成像控制系统产生的平面线扫描模式的控制波形,实现了高质量的人眼角膜的断层图像,验证了提出的设计方案的可行性。 In order to provide more portable and friendly human-machine interactivity control system for Optical coherence tomography(OCT),the control timing required by the frequency-domain OCT system in different scanning modes was analyzed.A design scheme of the embedded control system for human eye imaging was proposed.The control system is based on STM32F407ZGT6 chip of ARMv7-ME architecture and μC/OS-Ⅲ embedded operating system.The composition and the running process of embedded control system were described in detail in terms of hardware and software.By using the planar line scanning control waveform generated by the control system,a high-quality OCT image of human cornea was realized.The experimental result confirms the feasibility of the design scheme.
作者 秦嘉 高儒将 安林 Qin Jia;ao Rujiang;An Lin(Guangdong Weiren Meditech Co.,Ltd,Foshan 528000,China;Weiren Medical Care(Foshan)Co.,Ltd,Foshan 528251;School of Physics and Optoelectronic Engineering,Foshan University,Foshan 528000,China;Guangdong-Hong Kong-Macao Intelligent Micro-Nano Optoelectronic Technology Joint Laboratory,Foshan University,Foshan 528225,China)
出处 《现代科学仪器》 2022年第3期5-10,共6页 Modern Scientific Instruments
基金 广东省“珠江人才计划”引进创新创业团队项目(2019ZT08Y105) 佛山市香港科技大学产学研合作专项校企合作研发项目(FSUST21-HKUST10E) 国家自然科学基金(81771883) 国家自然科学基金(81801746) 国家特聘高层次人才青年项目 高校特色创新研究项目(2019XJZZ01) 粤港澳智能微纳光电技术联合实验室(2020B1212030010)。
关键词 光学相干断层扫描 成像控制系统 架构芯片 人机交互 嵌入式操作系统 Optical Coherence Tomography Imaging control system Architecture chip Human-machine interaction Embedded operating system
  • 相关文献

参考文献4

二级参考文献47

  • 1张锦红,彭蓉.帕金森病患者的眼部体征[J].中国临床康复,2005,9(33):122-123. 被引量:2
  • 2H H Gilgen,R P Salsthe,W Hodel.Submillimeter optical refleetometry[J].IEEE Lightwave Technology,1989,7(3):1225-1233.
  • 3A F Fercher.Eye-length measurement by interferometry with partially coherence light[J].Opt.Lett.,1988,13(3):186-188.
  • 4J G Fujimoto.Optical coherenee tomography[J].Science,1991,254(5035):1178-1181.
  • 5D X Hammer,R D Ferguson,N V Iftimia,et al.Advanced scanning methods with tracking optical coherence tomography[J].Optics Express,2005,13 (20):7937-7947.
  • 6A RTumlinsin,J K Barton.Endoscope-tip interferprocter for ultrahigh resolution frequency domain optical coherence tomography in mouse colon[J].Optics Express,2006,14(5):1878-1887.
  • 7A F Ferchera.Dispersion oompensation for optical coherence tomography depth-scan signals by a numerical technique[J].Lasserc Optics Communications,2002,204(3):67-74.
  • 8A F Ferchera.Numerical dispersion compensation for partial coherence interferometry and optical tomography[J].Optics Express,2001,9 (12):610-616.
  • 9COLSTON B W,DASILVA L B,EVERETT M J,et al.Dental OCT[J].Optics Express,1998,3(6):230-238.
  • 10COLSTON B W,MATTHEW J,EVERETT M J,et al.Imaging of Hard-and Soft-tissue Structure in the Oral Cavity by Optical Coherence Tomography[J].Applied Optics,1998,37(16):3582-3585.

共引文献87

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部