碳纤维治疗床在Monaco中的建模及剂量学影响 被引量:1

Modeling and dosimetric effect of carbon fiber couch in Monaco
下载PDF
导出
摘要 目的研究放疗中碳纤维治疗床对患者剂量分布的影响。方法测量射野点剂量得到碳纤维治疗床对射线的衰减,并依此在计划系统Monaco中进行治疗床的建模。选择2019年5月至12月于复旦大学附属中山医院接受放疗的5例鼻咽癌、5例直肠癌和5例椎体转移患者,在放疗计划系统Monaco中分析添加治疗床前后患者剂量分布的变化,并对这15例的剂量验证结果进行比较分析。结果通过比较Monaco计算与实测的治疗床衰减确定Monaco中的治疗床模型。添加该模型后,鼻咽癌、直肠癌和椎体转移病例的Gamma3mm/3%验证通过率平均增加0.32%、0.26%和2.06%。治疗床对靶区剂量影响最大的是椎体转移,衰减最大处达处方量的5%左右;其次是直肠癌,少量点达4%左右;对鼻咽癌的影响最小,衰减基本在3%之内。射线穿过治疗床可引起皮肤剂量的增加,对鼻咽癌,皮肤剂量增加最大约为处方量的8%;对部分直肠癌和椎体转移病例,皮肤剂量增加高达16%。结论治疗床会引起靶区剂量的减少和皮肤剂量的增加。 Objective To study the dose effect of carbon fiber couch on target and skin in radiotherapy.Methods The dose attenuation of the carbon fiber couch was obtained by measuring point dose of the beam fields.Then the couch was modeled in Monaco.Five patients with nasopharyngeal cancer,5 patients with rectal cancer and 5 patients with vertebral metastasis who received radiotherapy at Zhongshan Hospital Affiliated to Fudan University from May to December 2019 were selected to analyze the changes of dose distribution before and after the addition of couch in Monaco Planning System,and the dose verification results of these 15 patients were compared and analyzed.Results The model of couch was established by comparing the Monaco calculation with the measured couch attenuation.The addition of this model increased the Gamma 3 mm/3%verification pass rates by 0.32%,0.26%and 2.06%,on average,in patients with nasopharyngeal cancer,rectal cancer,and vertebral metastasis.The largest attenuation of the couch on the target dose was observed in cases of vertebral metastasis,which were about 5%of the prescription.The second was in cases of rectal cancer,a small number of about 4%of the prescription.The couch had the smallest impact on cases of nasopharyngeal cancer,with the attenuation basically within 3%of the prescription.The couch could also cause an increase on the skin dose,up to about 8%of the prescribed dose for cases of nasopharyngeal cancer,while for cases of rectal cancer and vertebral metastasis,the skin dose increase was as high as 16%of the prescription.Conclusions The couch can decrease the target dose and increase the skin dose.
作者 李婷婷 张玉洁 许六军 嵇卫星 张建英 Li Tingting;Zhang Yujie;Xu Liujun;Ji Weixing;Zhang Jianying(Department of Radiation Therapy,Zhongshan Hospital Affiliated to Fudan University,Shanghai 200032,China)
出处 《实用肿瘤杂志》 CAS 2022年第4期356-362,共7页 Journal of Practical Oncology
关键词 肿瘤 治疗床 靶区剂量 皮肤剂量 neoplasms couch target dose skin dose
  • 相关文献

参考文献7

二级参考文献50

  • 1C, erig LH, Niedbala M, Nyiri BJ. Dose perturbations by two cation fiber treatment couches and the ability of a commercial treatment planning system to predict these effects [ J ]. Med Phys, 2010,37:322-328.
  • 2Poppe B, Chofor N, Ruhmann A, et ai. The effect of a carben- fiber couch on the depth-dose curves and transmission properties for megavoltage photon beams [ J]. Strahlenther Onkol, 2007, 183:43-48.
  • 3Mihaylov IB, Cony P, Yan Y, et al. Modeling of carbon fiber couch attenuation properties with a commercial treatment planning system [ J]. Meal Phys,2008 ,35 :4982-4988.
  • 4Vanetti E, Nicolini G, Clivio A, et al. The impact of treatment couch modelling on RapidArc [ J]. Phys Med Biol,54:157-166.
  • 5Spezi E, Angelinia AL, Romani F, et al. Evaluating the influence of the Siemens IGRT carbon fiber tabletop in head and neck IMRT /-JT. Rad/other Oncol.2008.89:114-122.
  • 6胡逸民.靶区剂量的确定和对剂量准确性的要求[A]//胡逸民.肿瘤放射物理学[M].原子能出版社,1999:614.
  • 7Pulliam KB, Howell RM, Followill D, et al. The clinical impact of the couch top and rails on IMRT and arc therapy[J]. Phys Med Biol, 2006,56(23): 7435-7447.
  • 8Njeh CF, Parker J, Spurgin J, et al. A validation of carbon fiber imaging conch top modeling in two radiation therapy treatment planning systems: pbilips pinnacle3 and BrainLAB iPlan RT dose [J]. Radiat Oncol, 2012,7: 190-201.
  • 9Elekta C. Radiation oncology accessories catalog[M]. Stockholm: Elekta company, 2013 : 27 28.
  • 10McCormack S, Diffey J, Morgan A. The effect of gantry angle on megavoltage photon beam attenuation by a carbon fiber couch insert[J]. MedPhys, 2005,32(2): 483-487.

共引文献41

同被引文献7

引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部