为精确测量LED灯具的光通量,建立了积分球+高精度快速光谱辐射计测量光通量的模型,阐述积分球原理,自吸收修正系数的测量步骤和推导过程。选取一只标称功率为9W的普通照明用自镇流LED灯进行试验。试验结果表明,自吸收修正系数对LED灯具光通量测量结果影响较大,光通量与自吸收修正系数成正相关的关系。
In order to accurately measure the luminous flux of LED lamps, the model of integrating sphere + high-precision fast spectral radiometer is established. The principle of integrating sphere, the measurement steps and derivation process of self absorption correction coefficient are described. Select a self ballasting LED lamp with nominal power of 9W for general lighting for test. The test results show that the self absorption correction coefficient has a great influence on the measurement results of luminous flux of LED lamps, and the luminous flux is positively correlated with the self absorption correction coefficient.
作者
赵相泽
ZHAO Xiangze(Yunnan Electronic Information Products Inspection Institute,Yunnan 650031,China)
出处
《电子技术(上海)》
2022年第5期294-296,共3页
Electronic Technology
关键词
光通量
积分球
高精度快速光谱辐射计
LED灯具
自吸收修正系数
luminous flux
integrating sphere
high precision fast spectral radiometer
LED lamp
self absorption correction coefficient