为促进符合方法在核物理实验和辐射测量中的应用,根据脉冲波形的时间特点和波形前后沿的时间相关性,利用RS触发器置位构成的反符合电路,设计了便携式单道分析器,其对单一核素活度的测量误差小于±10%,工作稳定可靠,可用于脉冲幅度分析中。
In order to promote the application of conformity method in nuclear physics and radiation survey,according to the correlation of the characteristics of the time of impulse waveform and the time of front and back edges of the waveform,portable single track analyzer is designed through the utilization of anti-coincidence circuit of RS trigger setting component.For single nuclide activity,the measurement error is less than±10%,and the job stabilization is reliable,so it can be used in the impulse amplitude analysis.